background image

The chamber must be disconnected from the mains supply

(by unplugging it!)

before the glass covers are taken off for

cleaning. 

Cleaning the outer cover:

Simply lift up the outer cover to take it off for cleaning.

Cleaning the inner cover:

Carry out the following succession of steps to take off the

inner glass cover for cleaning:

–  Lift the outer cover up and off

–  Remove the spacer bolts fitted between the inner and

outer glass covers

–  Draw the connecting plug for the cover heating down and

out

–  Use the screwdriver provided to unscrew the side plate

–  Take out the flat plug for the pane heating (on the side

where the specimen is introduced)

–  Take off the inner cover

Removal and re-assembly of the heating wires: 

–  Disconnect the connecting plug between the two heating

segments

–  To linearly expand the heating wires, which are under

mechanical tension, use an external source of current to

electrically heat each of the heating wire segments, one

after the other, with 4.5 V/12 A. When the heating wires

have been slightly lengthened in this way, carefully take

out the track pair of a heating segment from the covering

plate of the cover. Remove the leaf springs between the

tracks and the glass plate.

Artificial radiation sources

On the left side of the chamber base there is an opening (20)

which serves for introducing artificial radiation sources. Use

the head of the screw to push the plate to the right. The ball

with pin which is located behind the plate can be turned

around with the aid of the pin until the opening towards the

inner area becomes visible. Now you can insert an artificial

radiation source like, for instance, the thorium source by

PHYWE a few centimeters into the inner area of the

chamber.

Notes

A: General

To avoid deposits on the inside surface of the inner glass

cover, never disconnect the chamber from the mains supply

nor switch it off at the main switch. This ensures that the

heating circuit for the glass cover heating remains active

even when the chamber itself - controlled by the timer - is

inactive at night or over weekends. In this way, condensation

of alcohol on the inside surface of the inner glass cover, and

the contamination resulting from this, are avoided to a great

extent.

Please read in detail the attached descriptin of the

thermostat.

B: Cleaning

Periodic cleaning of the glass covers with a commercially

available glass cleaning agent is recommended.

Important!

 

www.phywe.de, © All rights reserved

09046.90… 99/0709

4

Illumination

2

3

Отзывы: