3  FUNCTIONAL AND OPERATING ELEMENTS 

3.1 Operating elements 

The sensor has an on-button and two LEDs for indicating the 
Bluetooth and battery charge status. 
 

On-button 

Press  the  on-button  for  more  than  3  seconds  to  switch  the 
sensor on and off 
 

Bluetooth-LED  

Flashing red every 2 seconds 

Not connected 

Flashing green every 2 seconds 

Connected to the ter-
minal device 

Flashing green every 4 seconds 

Running measurement 

 

Battery charge LED 

Flashing red every 5 seconds 

Low battery 

 

3.2 Measurement inputs

 

On  the  front  side  of  the  sensor  there  is  a  BNC  connector  to 
which  the  supplied  potassium  ion-selective  electrode  (PK-2-
004) can be connected. 

4  NOTES ON OPERATION 

The  device  fulfils  all  of  the  technical  requirements  that  are 
compiled in current EC guidelines. The characteristics of this 
product qualify it for the CE mark. 
This instrument is only to be put into operation under special-
ist supervision in a controlled electromagnetic environment in 
research, educational and training facilities (schools, universi-
ties, institutes and laboratories). 
 The  individual  connecting  leads  are  each  not  to  be  longer 
than 2 m. 
The instrument can be so influenced by electrostatic charges 
and  other  electromagnetic  phenomena  (HF,  bursts,  indirect 
lightning  discharges) that  it  no  longer  works  within  the  given 
specifications. Carry out the following measures to reduce or 
eliminate  the  effect  of  such  disturbance:  Ensure  potential 
equalization at the PC (especially with Laptops). Use screen-
ing.  When  a  total  failure  of  the  instrument  occurs,  unplug  it 
and plug it back in again for a reset. 

5  HANDLING 

This  section  describes  the start-up  of the  sensor and  the  re-
cording  of  measurement  data.  Please  read this  section thor-
oughly in order to avoid failures or operating errors. 
 

5.1  Start-up 

Switch the sensor on by pressing the on-button for more than 
3  seconds.  The  Bluetooth  LED  lights  up  red.  Start  the  soft-
ware and select the sensor.

 

 

There  is  a  9-digit  code 
on  the  back  of the sen-
sor  (Fig.2).  The  last  4 
digits  of  the  code  cor-
responds  to  the  last 
four digits of the sensor 
name  in  the  software 
(Fig.3).  
 
 

 

 

 

 

 

 

 

 

 

 

Fig.2 

 

This  allows  an  exact 
assignment 

of 

the 

sensors  with  the  soft-
ware.  
 
 

 

 

 

 

 

 

Fig.3 

 
After  the  sensor  has  been selected  in  the software, the  LED 
flashes green to indicate a correct connection.  
If the sensor is switched on and not connected, it switches off 
automatically after 5 minutes. 
 
Connect  the  supplied  potassium  electrode  PK-2-004  to  the 
BNC connector of the sensor. 
 
 
 
 
 
 
 
 
 
 
 

Fig.4 

 

5.2 Preparation  
 
Important:  The  indicated  molar  masses  refer  to  sub-
stances in the anhydrous state! 
 

Preparation of the activation solution: 
1  M  KCl  (one  molar  potassium  chloride  solution),  2.5  M 
MgAc

2

  (two  and  a  half  molar  magnesium  acetate  solution) 

and H

2

0 deionised are required as stock solutions.   

The  concentration  of  the  activation  solution  should  be  1  mM 
KCl, 50 mM MgAc

2

. Due to the strong dilution of the electro-

lyte solution we recommend to prepare it in larger quantities. 
For the preparation of 1 litre electrolyte solution you need: 
 

 

979 ml H

2

0 deionised 

 

1 ml 1 M KCl 

 

20 ml 2.5 M MgAc

2

 

 

 

Remove  the  protective  cap  from  the  lower  end  of  the  elec-
trode. The electrode must be immersed in the activation solu-
tion  for  more  than  2  hours  before  use.  Make  sure  that  the 
lower part of the electrode does not rest on the bottom of the 
container, this may damage the electrode. 
Rinse the electrode in deionised water. 
The  electrode  can  be  stored  permanently  in  the  activation 
solution. 
 

5.3 Calibration 

Perform  a  2-point  calibration  using  measureAPP.  Use  350 
mg/l  for  the  1st  measuring  point  and  3500  mg/l  for  the  2nd 
measuring  point.  Use  a  magnetic  stirrer  with  uniform  stirring 
speed for uniform mixing. 
You  need  a strengthening  solution to  prepare the calibration 
solution and to carry out measurements. For the preparation 
of 1 litre of strengthening solution you need: 
 

 

1000 ml H

2

0 deionised 

 

186.38 g magnesium acetate 

 
 
 
 
 
 