background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

12620-09 and 12620-10 / 2715

 

The  arrow  keys 



  of  the  navigator  are  used  for  the 

navigation  within  the  menu.  Selections  must  be  confirmed 

with the confirmation key 

The following subchapters explain the various menu items of 
the  Cobra4  Mobile-Link 2  based  on  the  menu  structure  in 
Fig. 11. 

 

5.4.1  Sensor/Range (1.1) 

In  the case  of sensors  with  more  than  one  measuring chan-
nel, the measuring channel that is to be used can be selected 
by  way  of  ◄►  (e.g.  in  the  case  of  acceleration: 

  

  

,  or 

  

).  For  every  channel,  a  different  measuring  range  (

−2…2, 

−4…4 or −8…8) can be set. Then, the unit (e.g. g or m/s

2

) can 

be  specified  (Fig. 13).  Confirm  the  settings  with  the  confirma-

tion key 

. This will lead you back to the "Sensor" submenu. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5.4.2

 

Sensor > Setup (1.2)

 

In the case of some sensors, the "Sensor" menu includes the 
additional menu item "Setup". 

The  following  section  provides  an  overview  of  the  individual 
changes that are possible for these special sensors: 

 

Temperature compensation 

 

This setting option is available for the following Cobra4 Sen-
sor-Units: 

 

-

 

Cobra4 Sensor-Unit Chemistry (12630-00) 

-

 

Cobra4 Sensor-Unit pH (12631-00) 

-

 

Cobra4 Sensor-Unit Condu (12632-00) 

 

It  enables  the  automatic  temperature  compensation  of  a  pH 
or conductivity measuring channel, e.g. T(1), with regard to a 
temperature  input,  or  the  manual  input  of  a  known  tempera-
ture  with  the  purpose  of  compensation  (Fig. 15,  first  and 
second point in the menu). 

Apart  from  the  manual  adjustment  of  the  temperature  com-
pensation,  it  is  also  possible  to  enter  the  temperature  of  the 
solution. As far as the Cobra4 Sensor-Unit pH (12631-00) is 
concerned, the manual control can be switched on 

 or 

off 

 

via a switch and by way of ◄►. 

 

Temperature coefficient  

 

This setting option is available for the following Cobra4 Sen-
sor-Units: 

 

-

 

Cobra4 Sensor-Unit Condu (12632-00) 

 

Input of a temperature coefficient for application to a conduc-
tivity measuring channel (Fig. 15, third point).

 

 
 
 
 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig. 14: Menu structure of the Cobra4 Mobile-Link

 

Fig. 13: Display of the "Range" menu for the "Accelera-

tion" sensor in the Cobra4 Mobile-Link 

Range 
 
Menu

 

Channel

 

    ax 

 

 

 

Range

 

-

2…2

 

 

 

Unit

 

    g  

 

 

 

 

 

 

 

 

 

Menu > Sensor

 

Zero 

Setup 

Date/Time 

Battery status 

Device 

Information 

Large Display 

5.1 

5.2 

5.3 

5.4 

5.5 

Datalogger 

Setup 

Files 

2.1 

2.2 

Calibration 

One Point 

Two Point 

Reset 

4.1 

4.2 

4.3 

Sensor 

Range 

Setup 

1.1 

1.2 

Utilities 

 

Stopwatch 

Water level 

6.1 

6.2 

GPS 

6.3 

Only in case of sensors with setting options 

Only in case of Mobile-Link GPS (12620-20) 

Содержание 12620-09

Страница 1: ...nsor Setup 1 2 5 4 3 Datalogger Setup 2 1 5 4 4 Datalogger Files 2 2 5 4 5 Zero 3 5 4 6 Calibration 4 5 4 7 Setup Date Time 5 1 5 4 8 Setup Battery status 5 2 5 4 10 Setup Information 5 4 5 4 11 Setup...

Страница 2: ...supply unit 3 FUNCTIONAL AND OPERATING ELEMENTS 3 1 Operating elements on the top The Cobra4 Mobile Link 2 is equipped with various elements for the measurement data display operation and status indi...

Страница 3: ...nity of the unit After a complete shut down of the unit it must be switched off and on again by way of the ON OFF switch operating ele ment If the unit does not react keep the ON OFF switch pressed fo...

Страница 4: ...performed up to three times 5 3 Measurement data recording process A measurement can be started and stopped by pressing the confirmation key The measurement data recording proc ess is indicated by a...

Страница 5: ...The file name consists of the date and a running counter YYMMDDxx mli xx 00 99 Ensure that the date and time are set in the Cobra4 Mobile Link This can be checked and if needed changed in the Setup m...

Страница 6: ...Unit Conductivity 12632 00 It enables the automatic temperature compensation of a pH or conductivity measuring channel e g T 1 with regard to a temperature input or the manual input of a known temper...

Страница 7: ...e following Cobra4 Sen sor Units Cobra4 Sensor Unit Radioactivity 12665 00 It is used to change the counter tube voltage in order to measure the counter tube characteristics Gate time This setting opt...

Страница 8: ...imer Counter 12651 00 5 4 3 Datalogger Setup 2 1 The following changes are possible Selection of the single value measurement mode by way of a switch This ensures that only one value will be re corded...

Страница 9: ...u item enables the measurement values of the measuring channels to be manually reset to zero To do so the switches of the desired channels are set from off to on by way of the arrow keys Fig 20 You ca...

Страница 10: ...played in the header The automatic rotation of the device display can be switched on and off Use the confirmation key in order to confirm a new setting 5 4 10 Setup Information 5 4 Here you can find g...

Страница 11: ...es minutes and seconds This menu item can only be used in combination with a Mo bile Link 2 GPS 12620 20 and GPS activated See also chapter 5 4 9 Coordinates will not be displayed unless satellites ha...

Страница 12: ...f the battery decreases strongly it is possible to change the battery Please note that damage resulting from a battery change are not covered by the warranty You can send the Mobile Link to the repair...

Страница 13: ...Unit Energy 12656 00 Cobra4 Sensor Unit Forceplate 12661 00 Cobra4 Sensor Unit Radioactivity 12665 00 Cobra4 Sensor Unit Sound Level 12669 00 Cobra4 Sensor Unit Weather air pressure humidity temperatu...

Отзывы: