Service Manual PhysioFlow Lab1
14 / 20
Version of Monday, 10 July 2017
4.
Instructions, Problems
Troubleshooting
The orange lamp is permanently on or flashes during measurement:
please call your « after sales service »
for instructions (customer s33 3 72 82 50 00 or
, or your local dealer)
The device does not switch on:
please check fuses on back panel, and replace them if necessary. If problem
remains, then call your « after sales service » for instructions (customer support +33 3 72 82 50 00 or
, or your local dealer)
No ECG and no impedance signal are displayed:
verify that the patient cable is connected on PhysioFlow®
and that the serial connection is OK. Replace electric RS 232 serial link cable if the problem remains. If you
use an optical cable, check if the optical wire is properly plugged into the optical modules. You can as well
unplug the wire on both sides, (by pulling the locking tongues on both sides of the mode, and then pulling
the wire out of the module). Then use a sharp cutter, and make a neat cut at the wire extremity. Plug the
wire in the module again.
No ECG is displayed but other signals are visible:
verify the quality of electrodes and their position.
Electrodes must not be dry (use a bag that has been opened for less than one month).
ECG and Z signals are inverted:
launch again the calibration or acquisition phase in progress.