LIFEPAK 20/20e Defibrillator/Monitor Replacement
Procedures
Bottom Case
(continued)
W06 ECG Connector
Installation
To install the W06 ECG Connector:
1. On the outside of the case, align the W06 ECG Connector with the connector
standoffs and align the key in the connector with the notch in the bottom
case and slide the ECG connector into position.
2. Install two new 4-40
0.312 screws (173) into the W06 ECG Connector;
torque to 6.8 in-lb.
3. Position the cable in the slot between the first rib and the forward left corner
of the bottom case.
4. Place the first ferrite bead in its slot in the bottom case.
5.
6.
7.
.
8.
9. Complete the LIFEPAK 20 Device PIP.
9-137