OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Therapy cable damaged
•
Replace therapy cable and
perform daily checks per
Operator’s Checklist.
REPLACE BATTERY
prompt
occurs
Both batteries are very low
•
Replace one or both
batteries immediately.
•
Connect to auxiliary power
using approved power
adapter.
CPR time shown
(minutes/seconds) is different
than expected
Metronome is on
•
None. The metronome
adjusts the CPR time to
ensure CPR cycle ends with
compressions.
Incorrect setup option selected
•
Change CPR time setup
option. See
LIFEPAK 15 Monitor/Defibrill
ator Setup Options
provided
with your device.
Home Screen is blank but
ON
LED is illuminated
Screen not functioning properly
•
Print ECG strip to assess
rhythm and other active vital
signs.
•
Press
ANALYZE
and use
AED mode, if necessary.
For general troubleshooting tips, see General Troubleshooting Tips (on page 214).
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......