![Physik Instrumente C-863 Mercury Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/physik-instrumente/c-863-mercury/c-863-mercury_user-manual_1553908010.webp)
Introduction
CAUTION
!
Never connect a stepper motor to a C-863 DC motor controller. Irreparable
damage could result.
!
CAUTION
Do not enable a joystick axis here when no joystick is connected to the
controller
hardware. Otherwise the corresponding controller axis may start
moving and could damage your application setup.
!
CAUTION
The voltage output on the “Motor +” and “Motor -“ pins can be as high as the
supply voltage used. If using a non-standard power supply for the Mercury™
and a motor that connects to these lines (i.e. without separate PWM
amplifier), make sure the motor’s operating voltage will not be exceeded.
!
CAUTION
Never connect the RS-232-IN and USB connectors of the same controller to a
PC at the same time, as damage may result.
!
CAUTION—Maintain Compliance
Because the unit is not grounded over the power supply, a grounding screw is
provided at the lower left corner of the rear panel for connecting the metal
case to a protective ground. Connection to a protective ground is required for
compliance with applicable standards.
www.pi.ws
C-863 MS173E Release 1.2.7