Figure 3: Remote control software: control parameters 

7)

 

Setting of Pump-off Threshold: when input power is below the set Pump-off Threshold 
the Pump will automatically turn off. 

8)

 

Work Mode Setting 

 

ACC mode (constant current control mode): 0~400 mA 

 

APC mode (constant power control mode):   -8~17dBm 

Select desired working mode and input current/power value, then click Setting 
Button to apply.  

Click Read Button again to view the set value.  

 

5

 

Trouble Shooting 

 

Alarm

 

Possible Cause

 

Solution

 

Pump current alarm 

Internal problem 

Contact us for help  

Pump temp alarm 

Dirty or loose connector,  
or low input power 

Check input source 

Module temp alarm 

Internal problem 

Contact us for help 

High ambient temperature 

Ambient temperature needs to 
be within the specification 

Input alarm 

Dirty or loose connector,  
or low input power 

Check input source 

Output alarm 

High output  

Check input 

Internal problem 

Contact us for help 

Device disabled 

Enable device 

 

 

Содержание Agiltron EDFA3017B

Страница 1: ...1 15 Presidential Way Woburn MA 01801 Tel 781 935 1200 Fax 781 935 2040 www agiltron com C Band 17dBm Booster EDFA User Manual P N EDFA 1C2112332 S N 2021921079 November 19th 2021 A Photonwares Company ...

Страница 2: ... in C band with 17dBm maximum output Its front panel is shown in Figure 1 Figure 1 Front panel of EDFA The module needs 110V AC power Its on off power switch locates on its rear panel Note the RS 232 port on the rear panel is not functional for this model The FC PC connectors on its front panel are for laser inputs and outputs SMF 28 optical fiber cables are required for this model The micro USB p...

Страница 3: ...he Emission Button will be lighted up when output is on USB functions become active when Emission Button is ON 6 When turn off the EDFA ensure to deactivate power output first by pushing the Emission Button to light off and then turn off AC power 4 Software Instruction 1 Copy the GUI software from the USB flash drive coming with the device to host computer 2 Open the software Its interface is show...

Страница 4: ...ng Button to apply Click Read Button again to view the set value 5 Trouble Shooting Alarm Possible Cause Solution Pump current alarm Internal problem Contact us for help Pump temp alarm Dirty or loose connector or low input power Check input source Module temp alarm Internal problem Contact us for help High ambient temperature Ambient temperature needs to be within the specification Input alarm Di...

Отзывы: