15 Presidential Way Woburn, MA 01801 Tel: 781-935-1200 Fax: 781-935-2040
www.agiltron.com
C-Band 17dBm Booster EDFA
User Manual
(P/N: EDFA-1C2112332)
(S/N:
2021921079)
(November 19
th
, 2021)
A Photonwares Company
Страница 1: ...1 15 Presidential Way Woburn MA 01801 Tel 781 935 1200 Fax 781 935 2040 www agiltron com C Band 17dBm Booster EDFA User Manual P N EDFA 1C2112332 S N 2021921079 November 19th 2021 A Photonwares Company ...
Страница 2: ... in C band with 17dBm maximum output Its front panel is shown in Figure 1 Figure 1 Front panel of EDFA The module needs 110V AC power Its on off power switch locates on its rear panel Note the RS 232 port on the rear panel is not functional for this model The FC PC connectors on its front panel are for laser inputs and outputs SMF 28 optical fiber cables are required for this model The micro USB p...
Страница 3: ...he Emission Button will be lighted up when output is on USB functions become active when Emission Button is ON 6 When turn off the EDFA ensure to deactivate power output first by pushing the Emission Button to light off and then turn off AC power 4 Software Instruction 1 Copy the GUI software from the USB flash drive coming with the device to host computer 2 Open the software Its interface is show...
Страница 4: ...ng Button to apply Click Read Button again to view the set value 5 Trouble Shooting Alarm Possible Cause Solution Pump current alarm Internal problem Contact us for help Pump temp alarm Dirty or loose connector or low input power Check input source Module temp alarm Internal problem Contact us for help High ambient temperature Ambient temperature needs to be within the specification Input alarm Di...