Sampling and completion of Sanitization:
•
When you reach Sanitization Instruction Screen 5 you will start the 45 minute cycle by pressing
START CYCLE. The display shows the time remaining in the cycle.
•
As water enters the polishing filter housing it will begin to activate the ClCO
2
.
•
Install the supplied 3/8” barbed fitting into the dispense solenoid valve (where the POU filter was
removed)
•
6 feet of 3
/8” ID clear
vinyl tubing is provided in the kit. Connect the tubing to the barbed fitting.
Place the other end in a drain. NOTE: Insert the tubing directly into the drain, if possible, to
minimize odors.
WARNING
Chlorine dioxide gas is a respiratory irritant. Avoid breathing fumes. Ensure
area is sufficiently ventilated.
•
When 20 - 25 minutes remaining is displayed put on gloves provided in the kit and dispense a small
amount of water from the drain tubing into a small, clean container. Do not allow tubing to touch
the collected water. Replace the tubing into the drain.
•
Chlorine Dioxide test strips are provided in the kit. Use the white vial high range test strips to
confirm a minimum of 30 ppm chlorine in the sample by dipping a strip into the sample container.
Do not place the test strip directly under the running dispense water stream. Run water from the
sink when disposing of the water sample to minimize odor.
•
When the cycle ends, follow instructions to dispense 2
–
4 liters of water to drain. Wearing gloves,
take an additional water sample from the drain tubing into a small, clean container as above and use
the blue vial low range test strips to confirm <10 ppm of chlorine. If a greater amount is present
dispense an additional 2
–
4 liters and retest. Continue this process until <10 ppm is obtained.
•
Follow on-screen instructions to return system to normal operation.
NOTE: Dispose of the polishing filter by emptying into a sink while flushing with tap water or by
replacing the orange plugs and discarding. In this case no additional precautions are required. If
emptying, replace orange plugs before discarding.
Blue vial test strips for low range chlorine White vial test strips for high range chlorine
Содержание ACADEMY
Страница 1: ...ACADEMY CITATION BIOCHEM GENOME USER MANUAL...
Страница 6: ......
Страница 7: ...7 See Mounting Bracket Template...
Страница 36: ...Specification Sheet...
Страница 37: ......
Страница 40: ...40 DATE MAINTENANCE PERFORMED BY COMMENTS...
Страница 41: ...41 DATE MAINTENANCE PERFORMED BY COMMENTS...