40
41
Guidance and manufacturer’s declaration – emissions
The Innospire Mini is intended for use in the electromagnetic environment specified below. The customer or user of the
Innospire Mini should ensure that it is used in such an environment.
Emissions test
Group 1
RF emissions
CISPR 11
The Innospire Mini uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
Compliance
Electromagnetic environment – guidance
Class B
RF emissions
CISPR 11
Class A
Harmonics
IEC 61000-3-2
The Innospire Mini is suitable for
use in all establishments, including
domestic, and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Complies
Flicker
IEC61000-3-3
Guidance and manufacturer’s declaration – emissions
All ME equipment and ME systems
Guidance and manufacturer’s declaration – immunity
The Innospire Mini is intended for use in the electromagnetic environment specified below. The customer or user of the
Innospire Mini should ensure that it is used in such an environment.
Immunity test
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
ESD
IEC 61000-4-2
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the r/h should be
at least 30%
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
±2 kV mains
±2 kV I/Os
±2 kV mains
±2 kV I/Os
EFT
IEC 61000-4-4
Mains power quality should be that of a
typical commercial or hospital environment.
±1 kV differential
±2 kV common
±1 kV differential
±2 kV common
Surge
IEC 61000-4-5
Mains power quality should be that of a
typical commercial or hospital environment.
95% dip for
0.5 cycle
60% dip for
5 cycles
30% dip for
25 cycles
>95 % dip for
5 seconds
100% dip for
0.5 cycle
60% dip for
5 cycles
30% dip for
25 cycles
100% dip for
5 seconds
Voltage dips/dropout
IEC61000-4-11
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Innospire Mini requires
continued operation during power mains
interruptions, it is recommended that
the Innospire Mini be powered from an
uninterruptible power supply or battery.
3 A/m
3 A/m
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.