The device

Unplug the device and clean the external surface using one of the
following cleaning agents:
• A clean cloth dampened with water and a mild detergent
• 70% Isopropyl alcohol
• DisCide towelettes
• 10% chlorine bleach solution
Allow the device to dry completely before plugging in the power
cord. The device does not require routine servicing.

The breathing circuit, interface and bacteria filter

• After each use, the breathing hose and patient interface should be 

washed thoroughly with mild liquid detergent and water. The parts 
need to be completely dried before each use. 

• The bacteria filter cannot be washed. It should be replaced when 

blocked with mucus or trapped moisture.

• Replace the mask if it becomes torn or if you can no longer 

maintain a good seal.

The air filter

Under normal usage, you should clean the air filter at least once
every two weeks and replace it with a new one every six months.
• If the device is operating, stop the airflow. Disconnect the device 

from the power source.

• Remove the filter from the enclosure.
• Examine the filter for cleanliness and integrity. 
• Wash the filter in warm water with a mild detergent. Rinse 

thoroughly to remove all residue.

• Allow the filter to air dry completely. Replace the filter if it is torn 

or damaged. Only Philips Respironics filters should be used as 
replacement filters.

• Reinstall the filter.

Cleaning and maintenance

A patient guide to CoughAssist T70

8

A patient guide to CoughAssist T70

9

If the Information log picture      is displayed on the screen, it means information
is available in the information log screen but it does not necessarily affect therapy. 
To access these messages:
• From the Standby screen, press up (Menu) key, press down (Navigate) key 

to “Information Log.”

• Press right (Select) key. One of the following messages will appear:

Message

User action

Battery not charging – Temp

The detachable battery is too hot and cannot be charged. Allow device and/or 
battery to cool to resume charging. Remove battery and charge with battery charger 
accessory. If battery still cannot be charged, contact your homecare provider.

Check external battery

Power is being drawn from the detachable battery even though the external battery is 
usable. Replace external battery cable or external battery. If problem persists, contact 
your homecare provider.

Detachable battery not charging

The detachable battery cannot be charged. Replace battery. If problem persists with a 
different battery, contact your homecare provider.

Replace detachable battery

Detachable battery has failed or has reached end of life. Replace battery. If problem 
persists with a different battery, contact your homecare provider.

Internal fan failure – see User Manual

The internal fan is not working. Device should not be used with oxygen added to the 
patient circuit. Before using oxygen with this device, contact your homecare provider.

Card error

The device cannot write to or read from the SD card. Remove the SD card and use 
another card, if available. If problem persists, contact your homecare provider.

Information log

Содержание Respironics CoughAssist T70

Страница 1: ...es resulting from the use of this publication CAUTION US federal law restricts these devices to sale by or on the order of a physician Geyer JH 03 26 12 MCI 4104796 PN 1099607 Philips Healthcare is part of Royal Philips Electronics How to reach us www philips com healthcare healthcare philips com Asia 49 7031 463 2254 Europe Middle East Africa 49 7031 463 2254 Latin America 55 11 2125 0744 North A...

Страница 2: ...al and or mechanical assisted coughing technique A manual cough assist involves applying firm and rapid pressure to the upper abdomen forcing air out of the lungs A natural part of life CoughAssist T70 is a mechanical insufflation exsufflation device that does just that It simulates a natural cough by gradually delivering a large volume of air to your lungs similar to a normal deep breath when you...

Страница 3: ...on A patient guide to CoughAssist T70 4 A patient guide to CoughAssist T70 5 Product features 7 1 5 6 2 1 7 8 9 2 10 6 5 4 3 3 4 1 Fan exhaust Air from inside the device is expelled here 2 AC power inlet Connect the AC power cord here 3 Remote control connector Connect the optional remote control accessory foot pedal here 4 Connect a USB cable to this connector for service only 5 SpO2 connector If...

Страница 4: ...the set pressure Your device allows up to three prescription settings Your clinician will advise you on which prescription preset to use Getting started When using a mouthpiece nose clips may be required A well fitting soft seal mask is important to eliminate possible leakage Some patients may need to have their head supported from behind to prevent leaks In order to be effective firm pressure may...

Страница 5: ... patient guide to CoughAssist T70 9 If the Information log picture is displayed on the screen it means information is available in the information log screen but it does not necessarily affect therapy To access these messages From the Standby screen press up Menu key press down Navigate key to Information Log Press right Select key One of the following messages will appear Message User action Batt...

Страница 6: ...c Cough Trak n On n Off inspiratory trigger Inhale pressure ____ cm H2O Inhale flow n Low n Medium n High Inspiratory time ____ sec n NA Exhale pressure ____ cm H2O Expiratory time ____ sec n NA Pause time ____ sec n NA A patient guide to CoughAssist T70 10 A patient guide to CoughAssist T70 11 Optional accessories Description Item no Detachable battery 1043570 Detachable battery charger 1072182 D...

Отзывы: