Turning the Transceiver On/Off
4-4
Basic Operation
RF Auto
Shutoff
When RF Auto Shutoff is enabled (default), the transceiver stops broadcasting a
radio signal in order to prevent interference with other transceivers in use. This
condition occurs if there is no ECG signal for 10 minutes and the SpO
2
sensor
cable is
not
inserted in the SpO
2
sensor port. The technical alarm, "No Signal",
followed by the "Transmitter Off" INOP will be displayed at the Information
Center
.
Remove the batteries to conserve battery charge. However, if there is no
ECG signal for 10 minutes but the SpO
2
cable
is
inserted, the transceiver does
not stop broadcasting the signal.
To restart monitoring, insert batteries (if needed), attach leads to the patient,
insert the SpO
2
sensor cable in the SpO
2
sensor port and press the Check button
to verify association with the Information Center.
Standby
Mode
Standby mode is used to temporarily suspend monitoring at the Information
Center, for example, when a patient goes out of access point range or is
discharged.
Standby does not power down the transceiver, so if the standby period is
prolonged, you should remove the batteries.
Initiating
Standby
Step
Action
1
Do not remove the leads until after the patient is in Standby. This
will avoid an "ECG Leads Off" technical alarm, as well as
reminders, if configured.
2
In the Patient Window, select the
Standby
button.
3
Select the duration of the standby period (depending on your
system’s configuration: Infinite, 4 Hours, 3 Hours, 2 Hours, 1 Hour,
30 Minutes, 20 Minutes, or 10 Minutes).
Note—
If the patient will be discharged, select “Infinite” as the
standby duration.
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...