Telemetry Functions at the Information Center
9-18
Telemetry Functions at the Information Center & TeleMon
RF Auto
Shutoff
When RF Auto Shutoff is enabled (default), the transceiver stops broadcasting a
radio signal in order to prevent interference with other transceivers in use. This
condition occurs if there is no ECG signal for 10 minutes and the SpO
2
sensor
cable is
not
inserted in the SpO
2
sensor port. The technical alarm, "No Signal",
followed by the "Transmitter Off" INOP will be displayed at the Information
Center
.
Remove the batteries to conserve battery charge. However, if there is no
ECG signal for 10 minutes but the SpO
2
cable
is
inserted, the transceiver does
not stop broadcasting the signal. This feature can be disabled in Unit Settings.
Unit Settings - Alarms
Control
Function
Settings
Factory
Default
Non-
arrhythmia
Yellow
Alarms
Set latched/non-latched status
for SpO
2
, ST, and other non-
arrhythmia yellow alarms.
Latched
Non-latched
Latched
Leads Off
Adjust the severity level of this
technical alarm (INOP).
Cyan
Yellolw
Red
Cyan
Replace
Battery
Adjust the severity level of this
technical alarm (INOP).
Cyan
Yellow
Red
Cyan
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...