Draft - 1 Aug 08
ST/AR Arrhythmia Algorithm
ST/AR Arrhythmia Monitoring
7-5
ST/AR
Arrhythmia
Analysis
For information on arrhythmia detection, refer to the
following documentation:
•
ST/AR Algorithm - Arrhythmia Monitoring
Application Note,
#452298193051
•
IntelliVue Information Center Instructions for Use
and
Online Help
IntelliVue Telemetry does not have a dedicated cardiotach. Instead, the
arrhythmia cardiotach at the Information Center is used. Therefore, the ST/AR
Arrhythmia algorithm is always on for all IntelliVue Telemetry patients, and
cannot be turned off.
ST/AR provides Heart Rate and PVC Rate numerics and alarm detection for the
conditions listed in the following table. There are two detection levels: Basic and
Enhanced. Enhanced includes the Basic alarms.
Basic & Enhanced Arrhythmia Detection
Basic Arrhythmia Detection
Enhanced Arrhythmia Detection
Asystole
Non-Sustain VT
V-Fib/Tach
Vent Rhythm
V-Tach
Run PVCs
Brady yyy < xxx
Pair PVCs
Tachy yyy > xxx
Pause
HR yyy > xxx
Missed Beat
HR yyy < xxx
SVT
PVCs > xxx/min
R-on-T PVCs
Pacer Not Capturing
Vent Bigeminy
Pacer Not Pacing
Vent Trigeminy
Multiform PVCs
Afib
Irregular HR
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...