Troubleshooting
Maintenance, Troubleshooting, and Repair
7-71
Configuration settings
can be viewed by clicking on
Configuration
in the upper row
menu to display its menu, as shown in
Figure 7-39 HP NetRAID Assistant Configuration Menu
Clicking on
Display
, opens the
Raid System Configuration
window of
which shows the configuration settings of the NetRaid disks.
Figure 7-40 Raid System Configuration Window
Proper configuration settings for the 3 Raid disk drives are shown in
. If these
are not their configuration settings, RAID configurations may have to be restored. This
procedure is described in
LH3 - NetRaid Assistant Utility