HP8550
TECHNICAL INFORMATION
5. Stand
- The column of the stand has been mounted by means of
a hydraulic press and cannot be removed without causing
serious damage.
- As a result, the hinge housing cannot be disassembled
either.
- The legs have been mounted on the paw by means of a
hydraulic press and cannot be removed without causing
serious damage.
To replace the hinge housing assy:
- Undo the screws of the main grip.
- Take the main grip off stand.
- Remove the black plastic tube from the column.
- Slide the lamp housing together with the slider off the
column.
To replace the paw and the paw cap (L&R) and the wheels:
- Undo the screw on the inside of the paw.
- Unsnap the paw caps.
- The wheels are now accessible.
- Undo 4 screws in the point of rotation of the paw to detach
the paw.
To replace the sliding plug leg:
- To replace the sliding plug of the leg, the existing sliding
plug has to be smashed with the pointed end of a hammer;
otherwise it can’t be removed without damaging the leg.
A new sliding plug can simply be snapped onto the leg.
Always unplug the appliance before disassembly or repair!
1. Replacing lamps, fans and starters.
- Fold the appliance out into operating position.
- Remove 12 screws on the bottom of the lamp housing.
- Remove the upper part of the lamp housing
(the fans, starters and internal wiring are now accessible)
- Remove the screws of the mounting bridge
- Remove the reflector clamps.
- Remove the reflector.
You have to lift the mounting bridge to do so.
2. Replacing timer and lock grips.
- Remove the timer knob.
- Undo the 4 screws under the knobs.
- Remove the timer front.
- Undo 6 screws of the timer housing.
- Remove the front of the timer housing.
- Timer and lock grips are now accessible.
3. Replacing watchdog
- Start disassembling the appliance as described under 2.
- Undo the 4 remaining screws.
- Remove the back of the timer housing.
- The watchdog is now accessible.
4. Replacing seesaws (always replace both)
- Make sure the slider is in bottom position.
- Start disassembling the appliance as described under 2.
- Remove the lock grips.
- Remove spring. Be careful, for the spring is tensioned.
- Remove 2 shafts.
- Remove the seesaws.
DISASSEMBLY & REPAIR INSTRUCTION
- Input voltage
: 230 V – 50 Hz.
- Input consumption
: Approx. 2200 W
- Fuse
: 16 A
- UVA source
: 2 x Philips Cleo HPA SYNERGIE
- IR source
: 2 x Philips type 15040R
- Starter
: KAIDA KCCD-G400
- Timer
: 45 min. (motor driven)
- Radiation area
: 190 x 70 cm
- Output (min) UVX-36 meter
: 0.2 W/cm² at 65 cm. straight below the distance indicator
- Protecting goggles
: 2 x HB072
- Safety
: Class II Cenelec UV type 3
- Weight
: Approx. 19 kg.
2-8