Diagnostic Tests
2
-
22
Performance Verification and Safety Tests
ECG Tests
These instructions describe how to test the ECG functions.
Each of the ECG tests assumes the unit and the simulator are still set up as
they were at the end of the previous ECG test.
If all results are as described, the unit passes that portion of the test. Return to
the main Diagnostic Test menu by pressing
.
If there is any failure, begin troubleshooting and repairing the unit as needed.
See "Troubleshooting" on page 3-1 and Table 3-7 "Operational Problems -
ECG Monitoring (Pads or Leads)" on page 3-19.
1. Access the Diagnostic Test menu as described in "Diagnostic Mode" on
page 2-11.
2. Select
ECG Test
and press
.
3. The display should look similar to Figure 2-5:
Figure 2-5 ECG Test Display
Checking settings
In areas with 50 Hz power, the
AC Line Filter
should be set to
50 Hz
. If the set-
ting is incorrect, access the Configuration Mode and correct it. See "Configu-
ration Mode" on page 2-10.
The
PCI
setting should always be
Off
(factory set default). If it is
On
, press the
Volume Up and Down arrow keys simultaneously to turn it
Off
. (See Figure 2-
1 on page 2-11.) The software feature this setting controls is not used in the
M3500B/M5500B.
MAIN
ENTER
ECG TEST
Selected Lead:
Lead II
AC Line Filter:
60 Hz
On
Leads FE Status:
Good
Pads FE Status:
Good
DSP Status:
Good
DC Offset:
xxxx
Peak to Peak:
Diagnostic:
1050
Monitor:
1090
PCI:
51
Off
MAIN
ENTER
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...