22
G
uiDance
anD
m
anufacturer
’
S
D
eclaration
- e
lectromaGnetic
i
mmunity
: The goLITE BLU is intended for use in the electromagnetic
environment specified below. The user of the goLITE BLU should make sure it is used in such an environment.
i
mmuniTy
T
EsT
iEC 60601 T
EsT
l
EvEl
C
omplianCE
l
EvEl
E
lECTromagnETiC
E
nvironmEnT
-
g
uidanCE
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input-output lines
±2 kV for supply mains
±1 kV for input/output lines
Mains power quality should be
that of a typical home or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV for common mode
Mains power quality should be
that of a typical home or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
Mains power quality should be
that of a typical home or hospital
environment. If the user of the goLITE
BLU requires continued operation
during power mains interruptions, it
is recommended that the goLITE BLU
be powered from an uninterruptible
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical hospital
or home environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
EMC
Inf
or
ma
tion