4
Introduction
DreamStation BiPAP autoSV User Manual
Oxygen
When administering fixed-flow supplemental oxygen, the oxygen concentration
may not be constant. The inspired oxygen concentration will vary, depending on
the pressures, patient flows and circuit leak. Substantial leaks may reduce the
inspired oxygen concentration to less than the expected value. Appropriate patient
monitoring should be used, as medically indicated, such as an alarming pulse
oximeter.
When using oxygen with this system, the oxygen supply must comply with local
regulations for medical oxygen.
Do not connect the device to an unregulated or high pressure oxygen source.
When using oxygen with this system, a Philips Respironics pressure valve must be placed
in-line with the patient circuit between the device and the oxygen source. The pressure
valve helps to prevent the back flow of oxygen from the patient circuit into the device
when the unit is off. Failure to use the pressure valve could result in a fire hazard.
Oxygen supports combustion. Oxygen should not be used while smoking or in the
presence of an open flame.
Do not use the device in the presence of a flammable anaesthetic mixture in
combination with oxygen or air, in the presence of nitrous oxide, or in an oxygen-
enriched environment.
Do not use the device near a source of toxic or harmful vapors.
When using oxygen with this system, turn the device on before turning on the
oxygen. Turn the oxygen off before turning the device off. This will prevent oxygen
accumulation in the device.
Explanation of the Warning
: When the device is not
in operation and the oxygen flow is left on, oxygen delivered into the tubing may
accumulate within the device’s enclosure. Oxygen accumulated in the device
enclosure will create a risk of fire.
EMC
Medical Electrical Equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in this
manual. The device should not be used adjacent to or stacked with other equipment.
For more information, contact your home care service provider.
Portable and Mobile RF Communications Equipment can affect Medical Electrical
Equipment. See the EMC section of this manual for distances to observe between RF
Generators and the ventilator to avoid interference.
The Health Industry Manufacturers Association recommends that a minimum
separation of six inches be maintained between a wireless phone and a pacemaker
to avoid potential interference with the pacemaker. The DreamStation on-board
Bluetooth
communication should be considered a wireless phone in this regard.