Philips CM12 Скачать руководство пользователя страница 11

Erwin Sabio 

Philips CM12 Alignment & Setup Manual 

 Page 11 

Sample Loading & Unloading 
 

Mounting the Sample 

 

 

Figure 13 

An example of a TEM grid 

used to hold thin film samples

 

1.  Samples for the CM12 are deposited as thin films on a 

copper mesh grid such as in Figure 13.  This grid is the 
one loaded into the specimen holder (Figure 4) by first 
lifting the specimen clamp [

24

] using a pin-looking tool.  

(Note: Care must be taken to lift the clamp very gently 
so as not to break the fragile contraption.  In addition, 
never touch the rod between the o-ring [

23

] and the 

specimen cavity at any point during sample mounting.) 

 

   

 

2.   Using a pair of tweezers, carefully grab one side of the 

sample grid and lower it into the specimen cavity.  
(Note: Never grab the mesh in any other way so as not 
to distort the mesh or scrape away the sample.) 

 

   

 

3.  Gently lower the clamp in place using the pin tool to 

ensure the sample is locked in place. 

 

Inserting the Specimen Holder 
 

 

1.  Before inserting the specimen holder into the side entry 

chamber [

3

] (also called goniometer), check to make 

sure the tilt [

25

] is at 0 degrees and the lever [

28

] is in 

lock position. 

 

   

 

2.  Grab the end of the specimen holder nearest the black 

cap [

22

] and ensure that the slit on the cap is at the 8:00 

position when inserted to the chamber.  Push the holder 
further until a click as heard. 

 

   

 

3.  The red indicator light [

26

] will turn on as the chamber 

begins to be evacuated.  Wait for this light to turn off 
before proceeding to the next step. 

 

   

 

4.  When the light goes off, turn the specimen holder 

counter clockwise until the slit is at the 3:00 position. 

 

   

 

5.  Gently support the rod as it gets sucked further into the 

chamber to prevent damage to the tip with the sample. 

 

   

 

6. 

To “see” the sample on the large screen, spread the 
beam then adjust the stage controls [

8

] so that the 

viewing area is at the center of the opaque grid. 

 
 

Содержание CM12

Страница 1: ...Philips CM12 S TEM Alignment Setup Manual ...

Страница 2: ...Erwin Sabio Philips CM12 Alignment Setup Manual Page 2 ...

Страница 3: ...nt Column Alignment 9 Centering the Beam Alignment of Thermionic Gun Condenser Aperture Adjustment Condenser Astigmatism Correction Sample Loading and Unloading 11 Mounting the Specimen Inserting the Specimen Holder Sample Image Adjustment Optimizing Eucentric Height Adjusting Rotation Center Centering the Pivot Point Removing the Sample Obtaining the Image 14 Fine Tuning the Microscope Capturing ...

Страница 4: ...older High vacuum tunnel where the specimen is inserted 4 Objective Aperture Assembly An area that houses the objective aperture control knobs and related circuitry 5 Field Limiting Aperture Assembly An area that houses the intermediate lens aperture control knobs and related circuitry 6 Binoculars Eyepieces used to view the projected image up close 7 Viewing Window Glass windows another on the le...

Страница 5: ... Figure 3 The main control panel on the right hand side of the microscope 9 Intensity Knob 16 Emission Meter 10 Fine Adjustment Knob 17 High Tension Button 11 Screen 18 Filament Control Knob 12 Screen Control Buttons 19 Ready Button 13 Power Switch 20 Focus Focus Step Knob 14 Magnification Knob 21 Multifunction Control Knob 15 Beam Shift Knobs 9 10 11 12 16 17 4 18 15 19 20 21 13 14 ...

Страница 6: ...en Holder Figure 4 The CM12 Specimen Holder 22 Black Cap 24 Specimen Cavity Clamp 23 O ring The Side Entry Specimen Chamber Figure 5 The Side Entry Chamber 25 Tilt Knob 27 Z Adjustment Knob 26 Indicator Light 28 Tilt Lock Lever Startup Procedure 22 23 24 26 25 27 28 ...

Страница 7: ... be less than 35 5 Before powering the gun go to MODE TEM PARAMETER from the main screen to check the current and voltage setting that will be applied to the gun Ideally the current should be around 3 and the voltage at 100keV and should be kept at these levels Figure 7 The needle will kick twice to the right after the High Tension button is turned on 6 Finally push the HIGH TENSION 17 button to t...

Страница 8: ...green beam seen at the viewing screen represents visually the beam of electrons 4 As the filament saturation is being adjusted check the green beam on the large screen inside the viewing chamber 7 The brightness of this beam indicates the relative intensity of the actual electron beam Turn the filament knob to reach just enough brightness of the beam Turning the knob above the FIL LIMIT will not i...

Страница 9: ...rt by centering the cross over beam as described earlier Figure 9 The shadow of the filament becomes visible at low saturation 2 Undersaturate the gun by going to MODE CONFIGURATION and turning the Filament Knob 18 counter colckwise until the shadow of the filament becomes visible in the beam as in Figure 9 Figure 10 The STIG DF and ALGN buttons activate the multi function knobs 3 Press the ALGN b...

Страница 10: ...either direction past the cross over Condenser Astigmatism Correction Figure 12 Schematic of the symptom of condenser astigmatism 1 Asymmetric magnetic fields around the condenser lens results in an unbalanced focus of the beam This phenomenon is called astigmatism When this happens the beam spreads as asymmetric ellipses on either direction past the cross over when the intensity knob is turned St...

Страница 11: ...tly lower the clamp in place using the pin tool to ensure the sample is locked in place Inserting the Specimen Holder 1 Before inserting the specimen holder into the side entry chamber 3 also called goniometer check to make sure the tilt 25 is at 0 degrees and the lever 28 is in lock position 2 Grab the end of the specimen holder nearest the black cap 22 and ensure that the slit on the cap is at t...

Страница 12: ...rection Otherwise turn the other way 4 Continue alternating between turning the z knob and tilting the device until the movement of the reference point reaches a minimum At this point the optimal z height has been reached 5 Return the tilt to 0 degree then put the lever in lock position Sample Image Adjustment Adjusting Rotation Center 1 As in z height adjustment begin by choosing a central featur...

Страница 13: ... beam splits into two then the pivot point needs to be aligned Use the multifunction knobs to coincide the beams 3 Next select PIVOT PT Y and again coincide the beams if necessary using the multifunction knobs 4 Press the ALGN button again to exit the align mode Removing the Sample 1 Ensure once again that the tilt of the sample is at 0 and the level is in lock position 2 Pull the black cap of the...

Страница 14: ...ad the beam to illuminate a wide area and take out the viewing screen out of the way of the camera by pulling a lever on the left side of the viewing chamber 2 Switch to the computer and access the iTEM imaging software Click on the ACQUIRE button to see a live image of the sample on the computer screen 3 If ready click on SNAPSHOT to capture a still image of the sample 4 Click on STOP ACQUISITION...

Страница 15: ...ob counter clockwise until the filament level goes back to zero Note There is no wait time when decreasing the filament saturation 4 Turn off the HIGH TENSION switch and turn off the control panel If another user will use the microscope soon after just press VACUUM from the main menu to see the vacuum status then leave it at that 5 Cover the viewing windows with their lids 6 Logoff from the comput...

Страница 16: ...S 2007 January 19 230L first report Email attachment 6 Author s photo 7 Ibid 8 Ibid 9 Williams David B and Carter C Barry 1996 Transmission Electron Microscopy Volume I p 74 New York Plenum Press 10 Author s photo 11 Williams p 137 12 Ibid 13 Gilder Grids Retrieved 2007 January 17 from http www gildergrids co uk 14 Untitled Retrieved 2007 January 22 from http www kwc edu academic biology TEMpics t...

Отзывы: