ASSEMBLY TIPS + TEST CRITERIA AND CO POSSIBLE COMPLAINTS
ASSEMBLY TIPS
- Adjusting knob (11) ;
Before assembling the adjusting
knob (11), set the cam of the
adjusting unit 10 to the indicated
position.
- Adjusting the hand-strap
- Spring: Before replacing
TEST CRITERIA AND CONTROLS
Voltage
Ampere
Total product
13.0 V
380 mA ± 10 %
Adapter: Primary
voltage
220-240
V
Secondary voltage
13.0 V
400 mA
if < 12.5 V adapter is defective
Pump
13.0 V
180 mA ± 10%
Motor rollers
13.0 V
180 mA ± 10 %
N.B.
- To avoid dust from getting into the pump, the suction
opening has been closed with a plastic cap. Remove the
plastic cap before assembling.
- After the product has been repaired, it should
function properly and has to meet the safety
requirements as laid down and offi cially established
at this moment.
POSSIBLE COMPLAINTS
Status
Failure Cause
1. Switch in “OFF” position
Pump functions
ON/OFF switch defective
2. Switch set to level 1
Pump doesn’t function
Adapter defective
ON/OFF switch defective
Internal wiring broken
Defective
PCB
Motor pump defective
3. Switch set to level 1-5 and
LED burns
Filter is clogged
product not placed on the skin
Electronic switch defective
Defective
PCB
4. Switch set to level 1-5 and
Product doesn’t function
Electronic switch defective
product properly placed
Defective PCB
Sleeves got loose
Vacuum stop defective
Gearing motor defective
Pump clogged - fi lter missing
Internal wiring broken
To check, if the product functions properly.
- Remove sieve (13).
- Set the switch to level 1.
⇒
The rollers should rotate continuously and the LED doesn’t burn.
- Cover the suction opening with the thumb.
⇒
The LED lights up.
- Press the vacuum stop-buttons.
⇒
The LED goes out and the rollers stop rotating.
1
2
3
4
5
0
2
L
M
S
3
1
2
1
4
2 mm