Содержание burton SUPER EXAM 50 LED

Страница 1: ...ill t SUPER EXAM 50 LED FloorStand Table Rail Wall Ceiling Versions Instructions for Use Maintenance PHILIPS burton ...

Страница 2: ...5 SCHUKO SE50LR NEMA 5 15 SE50LRl NEMA 5 15 RAIL SE50LR02 BS 1363 100V 240V SE50LR102 BS 1363 SE50LR25 SCHUKO SE50LR125 SCHUKO SE50LW NEMA 5 15 WALL SE50LW02 BS 1363 SE50LW25 SCHUKO SE50LSC NEMA 5 15 CEILING SE50LSC02 BS 1363 SE50LSC25 SCHUKO RAIL has two different mounting brackets R Fairfield rail style Rl DIN 25mm rail style 5000249 A04 Instructions for Use SE50 LED 2 ...

Страница 3: ...blems that have not been treated in sufficient detail in this manual for your own safety please contact your supplier or Philips Burton before pro ceeding This manual contains guidance on how to use maintain the SE50 LED light Philips Burton is a leading manufacturer of medical lighting The product lines encom pass lights for examination procedure minor surgery and major surgery applications You a...

Страница 4: ... 4 ...

Страница 5: ...1 1 Pre Assembly 13 1 2 Wall or Table Models 14 1 3 Floorstand Models 15 1 4 Ceiling Models 15 1 5 Normal Operating Conditions 16 2 Maintenance 16 2 1 LED Replacement 16 2 2 Cleaning 16 2 3 Fuse Replacement 16 2 4 Replacement Parts 16 2 5 Preventive Maintenance 17 3 Symbols and Warnings 17 3 1 Symbol Explanations 17 3 2 Warnings and Cautionary Notices 18 Troubleshooting Guide 18 Warranty 19 Philip...

Страница 6: ...ht specifically de signed for use in the clinic or hospital setting Its design offers excellent illumination and color rendition large field size and convenient positioning SESO LED has a unique set of LEDs that gives the user a fixture that projects a tight homogeneous beam of light with high intensity to serve the needs of a demanding market Due to the low wattage the light can be in continuous ...

Страница 7: ...ght only when repairs or changes are carried out by Philips Burton or Philips Burton approved companies and components that influence the safety of the light are replaced with Philips Burton spare parts The light is not intended for use in areas where there is a risk of explosion This product complies with IEC EC UL 60601 1 for safety and IEC EN 60601 1 2 electromagnetic compatibility for medical ...

Страница 8: ...1060hPa Operation Ambient temperature 1O C to 40 C Relative humidity 30 to 75 Atmospheric pressure 700hPa to 1060hPa Shipping Container Markings Fragile Keep Dry Weights and Dimensions Net weight Iighthead flexible arm transformer housing cord 6Ibs 2 7 kg Net weight wall mount model 7 Ibs 3 2 kg Net weight floor model 28 Ibs 12 7 kg Lighthead widest diameter 6 25 in 15 9 cm Lighthead length 6 75 i...

Страница 9: ...II Diagrams U U 0 l u Q_ J r SE50 RAIL r NOTE All measurements are in inches SE50 FLOOR If 1 J 9 ...

Страница 10: ...Diagrams con t Wall Plate Mounting Pattern 10 353 8 j CV I I E9 t C Lf C e J 5 14 i This side up NOTE All measurements are in inches Floor Base Assembly ...

Страница 11: ...0 Ceiling Height i 3 85 __ 2 9 i 238 t 1 J I I 1 06 L NOTE All measurements are in inches I II 0 I 31 f r II I L CEILING HEIGHT H CLEARANCE DOWN TUBE 8 96 71 25 8 5 102 77 25 9 108 83 25 9 5 114 89 25 10 120 1 95 80 25 40 10 5 l26 86 40 11 132 97 40 11 ...

Страница 12: ...crews sIze M8 Note that one set screw has a dog point to engage a dimple on the tube 2 Slide the Mount Bracket up the tube and secure with the knob Actual height can be adjusted for the Mount Bracket 3 Attach the stop ring to the end of the drop tube and secure with set screw ...

Страница 13: ...ity if third party accessories are used Avoid spilling liquid on the lamp head NOTE Never open the transformer box while the power is connected After Installation is Complete After installation has been completed leave this manual for the end user 1 1 Pre Assembly 1 1 1 Remove the product and accessories from the shipping container s Save the containers and packaging in case hidden damage is revea...

Страница 14: ...ge This notation is to be made on the bill of lading 3 If in your opinion there may be concealed damage an agent from the freight carrier is obligated to make an inspection after the goods are unpacked 4 Do not destroy packing material until after the agent has made out his report 5 All claims must be made to the carrier not to Philips Burton 6 An RGA Returned Goods Authorization number must be ob...

Страница 15: ...shaft bushing into the upright 1 3 2 Floorstand Operation 1 3 2 1 Plug the integral power cord into a grounded convenience socket and activate the power switch The light should illuminate at full brightness 1 3 2 2 The lighthead can be positioned anywhere within its travel radius 1 4 Ceiling Models 1 4 1 Ceiling Assembly 1 4 1 1 Plan to install the ceiling mount preferably within 4 feet of an exis...

Страница 16: ... one or more system abnormalities do occur the protective current fuses s will fail and need to be replaced Before replacing a blown fuse s a complete system check should be under taken to reveal the underlying cause s for the failure Appropriate corrective action should be taken to prevent future occurrences 2 3 1 Turn the power switch off and unplug the fixture from the mains supply 2 3 2 Remove...

Страница 17: ... the mounting or C clamp for the table mount surface Annually Check the same items as for weekly and monthly maintenance If the answer to any of the questions is NO do not use the product Consult Inspect the power cord for external with your maintenance personnel abrasion before operating the light Lightly pull and twist the cord at the inlet strain relief to test for internal Any flickering is in...

Страница 18: ...e Cause Remedy No light Not plugged in Plug into suitable hospital grade receptacle output Switch not on Turn switch on Single or entire LED Have qualified personnel do the following board failed Contact Philips Philips Burton Fuses failed Replace with new fuse s Because of internal Locate short and fix it short Check condition of transformer and wiring Because of single occurrence of lightning st...

Страница 19: ...onform to the national or local technical safety standards in force in any country other than the ones for which the product was originally designed and manufactured 4 Lights for service under warranty must be either returned to the location instructed by your authorized Philips Burton representative or made available to the author ized Philips Burton field personnel where they will be replaced or...

Страница 20: ...91311 U S A Tel 800 444 9909 818 701 8700 Fax 800 765 1770 818 701 8725 Web www burtonmedical com EU Authorized Representative Philips Healthcare Nederlands BV PMS Quality Regulatory Affairs Veenpluis 4 6 5684 PC Best The Netherlands PHILIPS burton 20 ...