![Philips Brilliance C272P4 Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/philips/brilliance-c272p4/brilliance-c272p4_user-manual_1545064062.webp)
60
10. Troubleshooting & FAQs
Ans.:
Yes, there are 3 sizes to select: [
Small
],
[
Middle
], [
Large
]. You can press
to
enter OSD menu. Select your preferred
[
PIP Size
] option from the [
PIP / PBP
]
main menu.
Q3:
How to listen to Audio, independent
of video?
Ans.:
Normally the audio source is linked to
the main picture source. If you want to
change audio-source input (for example:
listen to your MP3 player independently
regardless of the video source input),
you can press
to enter OSD menu.
Select your preferred [
Audio Source
]
option from the [
Audio
] main menu.
Please note that the next time you
turn on your monitor, the monitor will
by default select the audio source you
chose last time. In case you want to
change it again, you need to go through
the above steps to select your new
preferred audio source, which then will
become the “default” mode.
+'0,
+'0,
$XGLR,Q
'LVSOD\3RUW
$XGLR6RXUFH
10.6 Clinical FAQ
Q1
:
Can I use color image in Clinical
D-image mode?
Ans.:
Clinical D-image mode is DICOM part-
14 calibrated for Grayscale performance
only.
Q2:
Can I clean the monitor and will it
affect its Antimicrobial housing ?
Ans.:
Yes, you can clean with normal monitor
cleaner or plain water. It will not affect
its Antimicrobial housing, as it is a
permanent part of the body (not a
coating).
Q3:
Can I use alcohol to clean the
monitor?
Ans.:
Alcohol should not be used to clean the
monitor as it has potential to damage
or deform the plastics and LCD screen
and its relevant coatings.
Q4:
Can I use the monitor in a patient
vicinity environment?
Ans.:
Yes, this monitor can be used in the
patient vicinity environments for it
complies with MOPP of ANSI/AAMI
ES60601-1.