Single breast pump
Double breast pump
Service life
500 hours
Mode of operation
Continuous operation
Non-continuous operation
55
min on, 180
min off at 22
°C
(72
°F) (duty cycle 23
%)
20
min on, 90
min off at 40
°C
(104
°F) (duty cycle 18
%)
Ingress protection
IP22 (motor unit), IP20 (adapter)
Explanation of symbols
The warning signs and symbols are essential to ensure that you use this device safely and correctly
and to protect you and others from injury. Below you find the meaning of the warning signs and
symbols on the label and in the user manual.
Symbol
Explanation
Indicates to follow the instructions for use.
Indicates important information such as warnings and cautions.
Indicates usage tips, additional information or a note.
Indicates the manufacturer.
Manufactured for: Philips Consumer Lifestyle BV, Tussendiepen 4, 9206 AD Drachten -
The Netherlands
Indicates the manufacturing date.
Indicates that the part of the appliance that comes into physical contact with the user
(also known as the applied part) is of type BF (Body Floating) according to IEC 60601-1.
The applied part is the breast pump kit.
Indicates 'Class II Equipment'. The adapter is double-insulated (Class II).
Indicates manufacturer's catalog number of the breast pump.
Indicates the batch number of the breast pump.
Indicates 'direct current'.
Indicates 'alternating current'.
28
English
Содержание Avent Premium SCF398, Avent Premium SCF396
Страница 1: ...Single Double electric breast pump Premium SCF398 SCF396 www philips com mybreastpump ...
Страница 2: ...Foldout left only ...
Страница 3: ...C B A A1 A2 A3 C5 C3 C4 C2 C1 A4 B1 B2 B3 B4 B5 B6 A5 ...
Страница 4: ...emtypageinsidecover ...