17 Monitoring Noninvasive Blood Pressure
148
Calibration
NBP is not user-calibrated. Cuff-pressure transducers must be verified at least once every two years by
a qualified service professional, and calibrated, if necessary. See the Service Guide for details.
Troubleshooting
Problem
Possible Causes
Solutions
Cuff will not inflate.
Monitor is in Service or Configuration
Mode.
Technical defect.
Call service.
Cuff tubing not connected.
Connect cuff tubing.
High or low values
measured (against clinical
expectations).
Contraction occurring.
Wait until contraction has finished.
Patient talking before or during
measurement.
Allow patient to rest quietly, then try again
after three to five minutes.
Incorrect cuff size or cuff not at heart
level.
Check cuff size, level, and position.
Noninvasive blood pressure reference
method set incorrectly.
Check the reference method configured
(auscultatory or intra-arterial) and correct if
necessary in Configuration Mode.
Displays zeros for systolic
and diastolic values.
Measurement automatically
repeats.
Severe vasoconstriction at cuff site.
Move cuff to another limb, check for shock,
or verify blood pressure using another
method.
Erratic blood pressure fluctuations due
to arrhythmias or rapid-acting drugs or
contractions.
Try again, if unsuccessful, verify blood
pressure using another method. Wait until
contraction has finished.
Excessive patient movement or
convulsions.
Restrain movement or verify blood pressure
using another method.
NBP CUFF OVERPRESS
INOP is displayed.
See “Patient Alarms and INOPs” on
page 73.
NBP EQUIP MALF
INOP is displayed.
NBP INTERRUPTED
INOP is displayed
NBP MEASURE FAILED
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...