Warnings and What To Do About Them

Base Station Warnings:

 

 blinks, alone or with lamps on either side. 

 means lamp is off/static on.

 means you hear an audible alert, if set. Press 

 to silence.

Transducer Warnings:

 

 blinks together with the symbol representing the problem source.

If you get this warning...

...Do this...

...On base station...

...On transducer...

Dock transducer to recharge battery, or replace transducer with a 
charged transducer.

If problem persists after a charge period of 30 minutes, change 
transducer battery. (See 

Service Guide

.)

Check that transducer is active and within range.

Check antenna connection.

Press 

 for two seconds, to release blocked slot.

Example bed label

This slot already has an 

active

 transducer (showing bed label 38 in 

this example). 

Either

 dock this one first to make the slot available

Or 

if the

 

transducer shows a bed label (16 in this example), return it 

to its own base station.

Example bed label

Example bed label

Call Support (transducer could not register).

If active ECG transducer, connect cable.

Place transducer in slot according to position indicated by the dot 
(transducer in wrong slot).

Base station and transducer Error Messages (E0 to E9):

 See 

Instructions for Use

M2720A Avalon CTS 

Cordless Fetal Transducer System

'$(& '