The following parameter settings are available in the AED Trainer 3. Default
settings are shown in
boldface
.
PARAMETER
SETTINGS
DESCRIPTION
Volume
loud
, medium,
soft
Sets the AED Trainer 3 speaker
volume.
Shock Series
1
, 2, 3, 4
Sets the number of shocks in a
series that must be delivered
before the AED Trainer 3
automatically activates a Basic
CPR protocol.
Shock Series
Interval (minutes)
1.0
, 2.0,
infinite
Sets the time interval used to
determine if a delivered shock
should be counted as part of the
current shock series. This
parameter applies only when the
shock series
setting is greater
than 1.
Metronome
on,
off
Enables (ON) and disables (OFF)
audio beats for CPR
compressions
AED Trainer 2
Remote Control
AEDT2 ,
AEDT3
Enables or disables use of the
AED Trainer 2 Remote Control
rather than the AED Trainer 3
Remote Control.
Adult Basic CPR
Duration
(minutes)
1.0, 1.5,
2.0
,
2.5, 3.0
Sets the length of the CPR
protocol for adult Basic CPR.
Infant/Child Basic
CPR Duration
(minutes)
1.0, 1.5,
2.0
,
2.5, 3.0
Sets the length of the CPR
protocol for infant/child Basic
CPR.
Содержание Aed Trainer 3
Страница 1: ...861467 Edition 2 A E D T r a i n e r 3 I N S T R U C T I O N S F O R U S E 453564382981 ...
Страница 6: ...P H I L I P S M E D I C A L S Y S TE M S INTENTIONALLY BLANK ...
Страница 32: ...P H I L I P S M E D I C A L S Y S TE M S INTENTIONALLY BLANK ...
Страница 33: ...P H I L I P S M E DI C A L S Y S TE M S INTENTIONALLY BLANK ...