IUP Quartz Transducer (1290C #J05)
Chapter 2 - Technical Specifications
DECG Transducer (M1357A)
Input Impedance:
>10M (di_erential, dc to 50/60Hz).
CMRR:
>110dB (with patient cable, 51.5k/0.047_F
imbalance atline frequency).
Noise:
<4_Vp (referred to input with 25k).
Contact Potential Tolerance:
_500mV.
Input Voltage Range:
20_Vp to 3mVp.
Patient Leakage Current:
<10_Arms @ 120V/60Hz.
Patient Auxiliary Current:
<0.1_A (dc).
Dielectric Strength:
2000Vrms (spark-gap protected).
Weight:
185g (6.5oz).
Cable Length
:
2.5m (8ft 2in) or 70cm (28in) for Option C03.
DECG/MECG Patient Module (M1364A)
The patient module has a7-pin ECG connector into which you can plug
either DECG cable (M1362A or M1362B) or MECG cable.
Overall length:
2706mm (+30, -100mm)
Length of free cable:
2618mm (+30, -100mm)
Weight:
120 grams
Size:
88x42x30mm
Socket:
DECG or MECG connection
A short cable version of the M1364A patient module with a 70 cm (28 inch)
cable is orderable as part number M1364-60003.
IUP Quartz Transducer (1290C #J05)
Dynamic Range:
-50 to 300mmHg.
Sensitivity:
5_V/V/mmHg.
Non-linearity:
whichever is greater:
±1% of the reading OR
±1mmHg.
Volume Displacement:
0.2mm3/100mmHg.
Weight:
180g (6.3oz).
Cable Length:
3m (9ft 10in).
Length:
37mm (1.5in).
Height:
21.7mm (0.9in).
Operating Temp:
15 to 40_C.
Содержание 50 T Series
Страница 4: ......
Страница 10: ...x List of Figures ...
Страница 26: ...Error Messages Chapter 3 Tests and Error Messages ...
Страница 35: ...Cardio Channel Chapter 4 Troubleshooting Cardio Channel Figure 7 Troubleshooting Cardio Channel ...
Страница 36: ...TOCO Channel Chapter 4 Troubleshooting TOCO Channel Figure 8 Troubleshooting TOCO Channel ...
Страница 39: ...Event Marker Chapter 4 Troubleshooting Event Marker Figure 11 Troubleshooting Event Marker ...
Страница 40: ...Nurse Call Chapter 4 Troubleshooting Nurse Call Figure 12 Troubleshooting Nurse Call ...
Страница 42: ... Chapter 5 Interfacing to a Fetal Monitor ...
Страница 48: ...Using the Service Program Chapter 6 Using the Service Software ...
Страница 51: ...Lists of Parts Chapter 7 Replacing Parts Figure 14 Transmitter Parts ...
Страница 53: ...Lists of Parts Chapter 7 Replacing Parts Figure 15 Receiver Boards ...
Страница 78: ...Receiver VCXO Test Chapter 8 RF Bench Repair ...
Страница 102: ...USA Chapter 11 Channels and Frequencies ...