39
150X3M LCD
Colour Adjustment
Go to cover page
13. Press or buttons to select R G B. Increase/decrease value by
press or buttons until the X, Y co-ordinates as below:
x= 0.281 0.005
y= 0.311 0.005
Y>= 200 nits
Fig. 5
0
6500 K
14. Setting X, Y value listed as below:
X= 0.312 0.005
Y= 0.338 0.005
Y>= 200 nits
Fig. 6
Alignment hits: 1. R for x value , G for y value, B for Y value on the
colour analyzer.
2. If the colour analyzer has been calibrated and preset
colour temperature in it. Please switch to correct
setting in accordance with colour settings.
Measurement/viewing selector
cover (black)
Fig. 1
Fig. 2
Fig. 3
0-CAL
4. Entering factory adjustment mode of LCD Monitor.
- To hold and buttons then power on the monitor. Press
to bring up OSD menu for confirmation.
Note : after alignment, please reset OSD to users mode for normal
operation. Otherwise, the monitor would not entering power
saving mode and showing full white picture all the time as no
video signal supplied. To leave factory mode by restart the
monitor.
5. Adjust OSD menu to lower position of screen (i.g. adjust V-position to
value 0 at submenu of OSD Setting.
6. Setting Brightness and Contrast
- Adjust Brightness to value100.
- Adjust Contrast to value 50.
7. Switch light probe to Viewing position.
8. Move the Lens barrel forward or backward to get clear image as
shown in Fig. 4
9. Switch light probe to Measuring position. It should be able to indicate
colour value on the CA-110.
Measurement/viewing selector
Fig. 4
Clear image
10. Setting pattern to full white picture.
11. Press then select hudson2 150 v2.15 Mar 27 2002 by button.
12. Press to bring up submenu as following windows.
Factory mode
HUDSON2 150 V2.15 Mar 27 2002
White Balance Adjustment
Alignment procedure
1. Turn on 150X3M LCD monitor.
2. Turn on the Timing/Pattern generator. See Fig. 1
Setting generator to provide CROSS-Hatch pattern at
Resolution :
1024 x 768
Timing :
H= 48 KHz
V= 60 Hz
3. Preset LCD colour Analyzer CA-110
- Remove the lens protective cover of probe CA-A30.
- Set Measuring/viewing selector to Measuring position for reset
analyzer. (Zero calibration) as Fig. 2
- Turn on the colour analyzer (CA-110).
- Press 0-CAL button to starting reset analyzer. See Fig. 3