47
SimplyGo Mini User Manual
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for Power Supply
Lines
±1 kV for Input/Output
Lines
±2 kV for Power Supply
Lines
±1 kV for Input/Output
Lines
Mains power quality should be that of a typical
home or hospital environment�
Electrical Transient
Conduction along
supply lines
ISO 7637-2
As specified in
ISO 7637-2
As specified in
ISO 7637-2
The Device is suitable for use in vehicles and
boats�
Surge
IEC 61000-4-5
±1 kV Line to Line
±2 kV Line to Ground
±1 kV Line to Line
±2 kV Line to Ground
Mains power quality should be that of a typical
home or hospital environment�
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
<5% U
T
(>95% Dip in
U
T
) for 0�5 Cycle at 45
degree increments
70 % U
T
(30% Dip in U
T
)
for 0�5 seconds
<5% U
T
(>95% Dip in U
T
)
for 5 Seconds
<5% U
T
(>95% Dip in U
T
)
for 0�5 Cycle at 45 degree
increments
70 % U
T
(30% Dip in U
T
) for
0�5 seconds
<5 % U
T
(>95% Dip in U
T
)
for 5 Seconds
Mains power quality should be that of a typical
home or hospital environment� If the user of the
Device required continued operation during
power mains interruptions, it is recommended
that the Device be powered from an
uninterruptible power supply or battery�
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical home or hospital environment�
Note: U
T
is the A�C� mains voltage prior to application of the test level�
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
SimplyGo Mini is intended for use in the electromagnetic environment specified below� The user of SimplyGo Mini should assure
that it is used in such an environment�
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should be used no closer
to any part of the Device, including cables, than the recommended 30 cm
separation distance�
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
150 kHz to 80
MHz
Radiated RF
IEC 61000-4-3
Radiated RF
IEC 61000-4-3
6 Vrms
Amateur
Radio &
ISM Bands
between 150
kHz and 80
MHz
10 V/m
80 MHz to
2�7 GHz
6 Vrms
Amateur Radio
& ISM Bands
between 150
kHz and 80 MHz
10 V/m
Interference may occur in the vicinity of equipment marked
with the following symbol:
AMSR
oxygenconcentratorstore.com
1-877-774-9271
Содержание SimplyGo Mini
Страница 3: ...AMSR oxygenconcentratorstore com 1 877 774 9271...
Страница 53: ...50 AMSR oxygenconcentratorstore com 1 877 774 9271...
Страница 55: ...52 SimplyGo Mini Manuel de l utilisateur AMSR oxygenconcentratorstore com 1 877 774 9271...
Страница 105: ...102 Manual del usuario del SimplyGo Mini AMSR oxygenconcentratorstore com 1 877 774 9271...
Страница 107: ...104 Manual del usuario del SimplyGo Mini AMSR oxygenconcentratorstore com 1 877 774 9271...
Страница 157: ...154 Manual del usuario del SimplyGo Mini AMSR oxygenconcentratorstore com 1 877 774 9271...