40
Procedure:
1. Place the dismantled nebulizers into the disinfector
basket:
A fixing of the nebulizers is only allowed, if they are
still totally movable in the mounting; the nebulizers
must not touch each other.
2. Place the disinfector basket with the nebulizers into
the disinfector (Attention: do not stack several disin-
fector baskets one on top of the other).
3. Start the program.
4. After the program has stopped, remove the disinfec-
tor basket with the nebulizers from the disinfector.
5. Assemble the nebulizers immediately upon removal
and pack them.
Recommended method:
manual procedure Cleaning
1. Place the dismantled nebulizers into the cleaning
solution for at least the time stated by the manufac-
turer of the cleansing agent / disinfectants.
2. Contamination on the outside can be carefully
removed with a soft brush.
3. Intensively rinse the nebulizers at least five times in
fresh, distilled or de-ionized water each; repeat this
cleaning procedure, if the last rinsing solution was
not clear or if there was still visible contamination on
the product.