Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

This Device is intended for use in the electromagnetic environment specified below. The user of the Device should assure that it is used in such 
an environment.

Immunity Test

IEC 60601 Test Level

Compliance Level

Electromagnetic Environment - Guidance

Portable and mobile RF communications equipment should be used no 
closer to any part of the Device, including cables, than the recommended 
separation distance calculated from the equation applicable to the 
frequency of the transmitter.

Recommended Separation Distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz

3 V

 

d = 1.2√P

Radiated RF
IEC 61000-4-3

3 V/m
80 MHz to 2.5 GHz

3 V/m

 d = 1.2√P     80 MHz to 800 MHz
 d = 2.3√P     800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts 
(W) according to the transmitter manufacturer and d is the recom-
mended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electro-
magnetic site survey, 

a

 should be less than the compliance level in each 

frequency range. 

b

Interference may occur in the vicinity of equipment marked with the 
following symbol: 

Note 1:  At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 

objects and people.

a.

Field strength from fixed transmitters such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, 
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment 
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which 
the Device is used exceeds the applicable RF compliance level above, the Device should be observed to verify normal operation. If abnor-
mal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Device.

b.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Содержание Actiwatch 2

Страница 1: ...Actiwatch Technician Guide...

Страница 2: ...wearer shows signs of skin reddening or inflammation If the device becomes damaged discontinue use and return it for replacement or repair Do not connect the device to a computer while worn or handle...

Страница 3: ...uide 10 pack Technician Clinician Guides Actiwatch 2 Communication Dock USB Wall Plug with Cable Actiware Software Replacement Watchbands andTool Wearer Guide 10 pack Technician Clinician Guides Symbo...

Страница 4: ...e Database New 3 For each new patient create a new patient record with a unique Subject ID in the database by selecting File New Subject 4 Connect your device to your computer with the provided USB ca...

Страница 5: ...Note Only Actiwatch 2 Spectrum PRO and Spectrum Plus devices are rechargeable Return Actiwatch Spectrum devices to Philips Respironics for battery replacement Actiwatch Spectrum PRO and Plus Connect y...

Страница 6: ...watch is five years WEEE Recycling Directives Separate collection for electrical and electronic equipment per EC Directive 2002 96 EC Dispose of this device in accordance with local regulations Contac...

Страница 7: ...ts for Safety IEC 60601 1 2 2007 Medical Electrical Equipment Part 1 2 General Requirements for Safety Collateral Standard Electromechanical Compatibility Requirements andTests IEC 60601 1 11 2010 Med...

Страница 8: ...magnetic environment specified below The user of the Device should assure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidanc...

Страница 9: ...s determined by an electro magnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbo...

Страница 10: ...nsmitter Watts Separation Distance According to Frequency of Transmitter meters 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1...

Страница 11: ...s or workmanship This warranty does not cover damage that may occur in shipment This warranty does not apply to any Product or individual parts that have been repaired or altered by anyone other than...

Страница 12: ...1104768 1104715 R01 JDH 04 24 2013...

Отзывы:

Похожие инструкции для Actiwatch 2