2-1
SECTION 2: TECHNICAL SPECIFICATIONS
AC/DC KILOVOLTMETERKVM50E
KVM50 DIVIDER
HIGH VOLTAGE AC INPUT:
50kVAC maximum
HIGH VOLTAGE DC INPUT:
50kVDC maximum
HIGH VOLTAGE PK INPUT:
74kV PEAK (AC)
DIVIDER RESISTANCE:
300M
DIVIDER CAPACITANCE:
≤ 15 pF
DIVIDER RATIO:
10,000:1
INSTRUMENTATION:
AC / BATTERY POWERED
POWER ADAPTOR:
INPUT: 85-264 Volts AC, 0.6 AMPS, 47-63 HZ
OUTPUT: +15 VDC, 1.66 AMP
INPUT JACK:
+15 VDC, 1 AMP
BATTERY PACK:
9.6 Volt Ni-MH 3800 mA hr
VOLTAGE MEASUREMENT:
Voltage is measured by means of Peak Responding, RMS or AVG
Circuitry. 4 ½ Digit LCD display.
RANGES:
Low Range
– 20kV AC & DC
High Range
– 50kV AC & DC
ACCURACY: 1% of Reading from 10% -100% of Range
FUNCTION SELECTIONS:
1. Peak /
2 (Peak responding calibrated to RMS Value)
2. Peak (Peak responding calibrated to Peak Value)
3. RMS (True RMS)
4. AVG (True Average Value,
not
RMS equivalent)
5. Peak DC (Peak responding including ripple peak
6. Ripple Voltage (RMS value of a DC Ripple voltage)
7. DC RIPPLE (Peak to peak AC ripple on DC voltage / 2)
FREQUENCY RESPONSE:
(SINUSOIDAL WAVEFORM)
AC and Ripple voltage measurement functions in the range of 20-
1000Hz.
SETTLING TIME:
AC Peak, AC Peak /
2,
DC Peak, DC Ripple
AC AVG, AC RMS, DC AVG
--Up to 30 seconds
--Up to 5 seconds
ENVIRONMENTAL
CONDITIONS:
10-40°C, Indoor/Outdoor in fair weather
Humidity <95%, non-condensing
Altitude <3000 ft (1000 meters)