SECTION 4:  ELECTRICAL SET-UP / OPERATIONAL NOTES 

 

 

 

Reference figure 1. 
 
1. 

Place test set in desired location.  
 

2. 

Connect a ground of sufficient size from source ground to grounding post located at base of the 
divider (

10’ lead is supplied with unit).   

 

3. 

Connect supplied coaxial cable between low voltage arm, located at the base of the divider, and 
metering module (

25’ lead is supplied with unit).   

 

4. 

Connect ground post on metering module to source 

ground (10’ lead is supplied with the unit).   

 
5. 

Connect source voltage to be measured to top of divider by an appropriate and safe method. High 
voltage should enter divider from above at an angle of no greater than forty-five degrees from 
vertical for highest accuracy readings. 

  
 

CAUTION:  Metering module and divider base must be grounded for safe and proper operation. 

Damage to unit mat result if left ungrounded. 

 

NOTE:  Unit is designed to measure voltage in reference to ground.  Source voltage and divider 

ground post must be ground referenced for unit to operate properly.   

 
NOTE:  Coaxial lead is part of calibration. Significantly changing length of meter lead will affect 

AC calibration. 

 

 

WARNING! 

THIS UNIT SHOULD ONLY BE OPERATED BY PERSONS KNOWLEDGEABLE OF HIGH 
VOLTAGE TESTING AND SAFETY PROCEDURES.  IMPROPER OPERATION MAY RESULT IN 
INJURY OR DEATH.

  

 

 

ENSURE THAT UNIT TO BE TESTED IS DE-ENERGIZED AND DISCHARGED!  ENSURE THAT 
WORKING ENVIRONMENT IS SAFE AND FREE OF HAZARDS.   

Содержание KVM 100A

Страница 1: ...NUAL AC DC KILO VOLTMETER Model Number KVM 100A 200A PHENIX TECHNOLOGIES INC 75 Speicher Drive Accident MD 21520 Phone 1 301 746 8118 Fax 1 301 895 5570 E mail info phenixtech com Copyright Phenix Tec...

Страница 2: ...NICAL SET UP 3 ELECTRICAL SET UP OPERATIONAL NOTES 4 OPERATING INSTRUCTIONS 5 CALIBRATION 6 MECHANICAL MAINTENANCE 7 STORAGE OF EQUIPMENT 8 TROUBLESHOOTING 9 PARTS LISTS 10 RECOMMENDED SPARE PARTS 11...

Страница 3: ...KVM100A 200A 1 1 DANGER WARNINGS Complete Grounding of this unit is necessary for the safe operation of this equipment Disconnect inputs before ungrounding this equipment...

Страница 4: ...ed below for your reference Become familiar with your instrument before performing an actual test Know your work area check that all circuits are de energized and locked out Never work alone always wo...

Страница 5: ...Range FUNCTION SELECTIONS 1 AC AVERAGE Average absolute value not RMS equivalent 2 AC RMS True AC RMS 3 AC PEAK Average of the positive and negative peak values 4 AC PEAK 2 AC PEAK 2 5 DC AVERAGE Ave...

Страница 6: ...to determine if the unit was damaged in shipment If there are any signs of physical damage such as dents scratches or oil leaks contact the Service Department at Phenix Technologies before proceeding...

Страница 7: ...ter divider from above at an angle of no greater than forty five degrees from vertical for highest accuracy readings CAUTION Metering module and divider base must be grounded for safe and proper opera...

Страница 8: ...KVM100A 200A 4 2 ELECTRICAL SET UP OPERATIONAL NOTES Figure 1 KVM Testing Setup...

Страница 9: ...ttery charging or AC operation with included power adapter 5 GROUND STUD Ground stud for connecting metering module to source ground 6 BNC INPUT CONNECTOR BNC connector to connect the metering module...

Страница 10: ...it will not over range on the AVG or the DC modes before attempting to read the ripple voltages DO NOT attempt to read ripple voltage on a range lower than required to measure the full input voltage 5...

Страница 11: ...eadings will correspond to the type of meter used AVG will show AVG RMS will show RMS etc Use DC range for DC signals 10 meg input meter 3 Select desired mode for voltmeter 4 Activate high voltage sou...

Страница 12: ...S It is recommended calibration be performed on yearly basis Due to the complexity and the standards required to calibrate this instrument Phenix Technologies does not recommend customer calibration F...

Страница 13: ...t the elements in normal use It is recommended that finish be wiped down with nothing stronger than ordinary household cleaner for longer life and for proper electrical operation of the unit Also insp...

Страница 14: ...r a prolonged period the following precautions are recommended Equipment should be covered and kept in warm dry environment 95 maximum humidity 5 to 50 degrees C Prior to placing equipment back into o...

Страница 15: ...n Battery may be discharged Recharge battery Unit blanked out during testing because of transient discharge occurrence and won t come back on by cycling the power switch Carefully remove front panel a...

Страница 16: ...D PANEL 1 NE1 1609990 NEON LAMP 1 PCB 1372 31137200 BATTERY POWER MANAGER 1 PCB 1405 31140500 INTERFACE BRD 1 PCB 1417 31141700 METERING BRD 1 PS1 1590050 POWER SUPPLY 1 PS1 1077167 POWER CABLE DIVIDE...

Страница 17: ...ormal operation If the unit will be operated at an isolated site for an extended period or will be subjected to unusual stresses a larger quantity of parts should be stocked as spares In such a case c...

Страница 18: ...to include the serial number of the unit as well as the part number of the replacement part When your purchase order is received at our office a representative of Phenix Technologies will contact you...

Страница 19: ...umber Reason for Return Cause of Defect If Phenix Technologies Inc deems return of the part appropriate it will then issue an Authorization for Return If return is not deemed advisable other inspectio...

Страница 20: ...Z D TR M K K K TB X Intensite Diode Transisteur Thyristor Borne Ouvert Ferme Contact de SYMBOLE ZU SCHEMA SYMBOLES POUR SCHEMA DE CIRCUIT CIRCUIT DIAGRAM SYMBOLS Variable Resisitor Resistor Lamp Indi...

Страница 21: ...KVM100A 200A 15 1 SECTION 15 SCHEMATICS AND DRAWINGS Drawing Number Description 1 7916005 KVM100A 200A SCHEMATIC...

Страница 22: ...KVM100A 200A...

Страница 23: ...rator s Manual are correct If there are concerns or comments as you have used this information Phenix Technologies appreciates any feedback Unit Serial Number Sect Page s Comment Please return to Phen...

Отзывы: