16
SETTING OF ALARM RESUME TIME
The buzzer and remote alarm are cancelled by pressing buzzer stop key (BUZZER) on the control panel
during alarm condition. However, if the alarm condition is continued after the “alarm resume time” has
passed, the buzzer sounds again.
It is possible to change the alarm resume time by the procedure shown in the Table.5.
The example in the table is based on the assumption that the desired duration is 20 minutes.
Note:
Factory setting; 30 minutes.
Table 5. Setting procedure for alarm resume time (change from 30 minutes to 20 minutes)
Description of operation
Key operated
Indication after operation
1
----
The current chamber temperature is
displayed.
2 Press digit shift key for 5 seconds.
The first digit is flashed.
3
Set the figure to F25 with the digit
shift key and numerical value shift
key.
The settable digit is shifted.
When pressed, the figure of settable
digit changes.
4 Press set key.
SET
The current resume time is displayed.
The second digit is flashed.
5
Set the figure to 020 with the
numerical value shift key.
When pressed, the figure of settable
digit changes.
6 Press set key.
SET
The setting is memorized and
the current chamber temperature is
displayed.
• The settable alarm resume time are 10, 20, 30, 40, 50, or 60 minutes. The buzzer and remote alarm
would not resume if the resume time is set in 000.
• The setting of alarm resume time cannot be changed during the defrosting.
• The buzzer and remote alarm during power failure or alarm testing cannot be silenced.
• The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.