37
OPERATION/ALARM LOG
7.
When the export is complete, Information
dialog box is displayed. Press OK key.
Note:
Even after the export of operation log
data is complete, operation log data saved in
the unit are not deleted.
8.
Remove the USB memory from the USB port.
Note:
• The log folder is created in the USB memory, and the exported file is saved in it in CSV format. The
exported file name is in date (8 digits) - type of data format.
(e.g.) When exporting all types of data using All (from Jan. 1st, 2019 to Oct. 1st, 2019):
20190101-20191001_AllCh.csv
20190101-20191001_Door.csv
(e.g.) When exporting Actual Temp. using 1 Day (Jan. 1st, 2019):
20190101_Temp.csv
If the file name is duplicated, A sequential number such as "-1" is added to the end of the file name to
be output later.
• On the beginning of the exported file, product name (MDF-DU901VH) is written. However when the
Unique ID is registered [page 32], product name and Unique ID (8-digit) are written.
(e.g.) When “RoomA001” is set as the Unique ID of MDF-DU901VHA:
MDF-DU901VH, RoomA001
9.
Press Top key to return to the Top screen.