   ROUTINE MAINTENANCE 

 

    Defrosting of inside wall

 

 
The frost is built at the upper portion of the chamber and inner door.    The excessive frost possibly make 
some gap between the cabinet and the magnetic door gasket, which may cause poor cooling.    Remove 
the frost on the inner door with a scraper enclosed with the unit.  Following shows the procedure for 
removing the chamber frost. 

Note:

    For removing the frost, do not use a tool with sharp edge such as a knife or a screw driver. 

 
1. 

Turn off the backup cooling kit if applicable.

 

 
2.

 Take out and transfer all the contents to another freezer or a container which is refrigerated by liquid 

carbon dioxide or dry ice. 

 
3. 

Turn off the power switch and battery switch of the 

freezer.

 

 
4.

 Open the outer door and inner door.  Remove the 

inner door by lifting up as shown in the figure.   

 
5.

 Leave the freezer as it is. 

 

6.

 The water accumulated on the bottom of the 

chamber should be wiped up with a dry cloth. 

 
7. 

After cleaning the chamber and inner door, replace 

the inner door and start up the unit according to the 
procedure on page 14. 

 
8.

 Put back the articles into the sufficiently cooled 

freezer compartment. 

 
9. 

Turn on the backup cooling kit if it is provided. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Содержание MDF-DU500VH

Страница 1: ...H Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 42 for model number Operating Instructions Ultra Low Temperatur...

Страница 2: ......

Страница 3: ...ONITOR OF FREEZER STATUS P 21 ALARMS SAFETY FUNCTIONS P 22 ROUTINE MAINTENANCE P 23 Cleaning of cabinet P 23 Cleaning of condenser filter P 23 Defrosting of inside wall P 24 Cleaning of air intake por...

Страница 4: ...ating instructions are subject to change without notice for improvement of performance or function Contact our sales representative or agent if any page of the operating instructions is lost or the pa...

Страница 5: ...Symbol shows This symbol means caution This symbol means an action is prohibited This symbol means an instruction must be followed Be sure to keep the operating instructions in a place accessible to u...

Страница 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 7: ...supply plug not the cord Pulling the cord may result in electric shock or fire by short circuit Disconnect the power supply plug before moving the unit Take care not to damage the power cord A damaged...

Страница 8: ...insertion may cause a heat or ignition Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components or electric...

Страница 9: ...terior Flammable refrigerant label Sample damage Chamber temperature Interior Forgets to close a door and latch Sample damage Chamber temperature Interior Rise in temperature is prevented Personal inj...

Страница 10: ...f OVERVOLTAGE CATEGORY II Temporary OVERVOLTAGES occurring on the mains supply Applicable pollution degree of the intended environment POLUTION DEGREE 2 in most cases INTENDED USE AND PRECAUTIONS This...

Страница 11: ...9 FREEZER COMPONENTS Rear side 11 inside 7 inside 1 2 3 6 8 12 7 4 9 10 13 5 17 18 16 15 Power cord 14 MDF DU700VH...

Страница 12: ...ided to facilitate moving of the cabinet For the installation adjust the leveling feet so that the front 2 casters cannot contact with the floor 11 Condenser filter behind the grille This filter preve...

Страница 13: ...seconds in the temperature display mode leads setting mode for alarm temperature alarm resume time compressor delay time and door alarm delay time Refer to page 16 17 18 and 19 for details respective...

Страница 14: ...on or noise Unstable condition may cause vibration or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in inj...

Страница 15: ...me to the wall with these fixtures and rope or chain Fig 2 4 Setting the shelves 3 shelves are packaged at the bottom of the chamber Set the shelves on the shelf stoppers at the standard location firm...

Страница 16: ...RM TEST with the battery switch off 8 After confirming the above you can put articles into the chamber in a small batch to prevent the temperature rise 9 Push the test switch to check the operation of...

Страница 17: ...ature of 30o C The setting mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation auto return function In this case the setting is not accepted...

Страница 18: ...the figure to 005 by using digit shift key and numerical value shift key When pressed the figure of settable digit changes When pressed the changeable digit is shifted 6 Press set key SET Alarm temper...

Страница 19: ...ure to F25 with the digit shift key and numerical value shift key The settable digit is shifted When pressed the figure of settable digit changes 4 Press set key SET The current resume time is display...

Страница 20: ...igit is flashed 3 Set the figure to F05 with the numerical value shift key When pressed the figure of settable digit changes 4 Press set key SET The current delay time is displayed The first digit is...

Страница 21: ...ss numerical value shift key for 5 seconds The first digit is flashed 3 Set the figure to F04 with the numerical value shift key When pressed the figure of settable digit changes 4 Press set key SET T...

Страница 22: ...when the outer door is closed and opened soon 1 Turn the cap on the left side counterclockwise about two laps Or remove the cap 2 Allow about twenty seconds before open the outer door 3 Close or repla...

Страница 23: ...mance or durability of refrigerating circuit Examination is air conditioning of installation site Notice of overload condition When the running rate of refrigerating circuit is higher than usual Statu...

Страница 24: ...r temp are displayed alternately Intermittent tone Remote alarm If the cascade sensor is short circuited Alarm lamp is flashed E04 and chamber temp are displayed alternately Intermittent tone Remote a...

Страница 25: ...oing so can damage the electric insulation and cause failure The compressor and other mechanical part are completely sealed This unit requires absolutely no lubrication Note Always replace the inner a...

Страница 26: ...up cooling kit if applicable 2 Take out and transfer all the contents to another freezer or a container which is refrigerated by liquid carbon dioxide or dry ice 3 Turn off the power switch and batter...

Страница 27: ...ke port cleaning The outer door cannot be opened even if the cap on the air intake port is removed Remove the frost and ice with the stick for air intake port cleaning The frost and ice are built in t...

Страница 28: ...situation Important The used battery is a recyclable precious resource Do not dispose of the battery Always follow the procedure for recycling WARNING The replacement of the battery for power failure...

Страница 29: ...ed The condenser filter is clogged Always clean the condenser filter when the filter check lamp is lit The outer door is not shut tightly The inner door is not installed correctly The set temperature...

Страница 30: ...is product contains flammable refrigerant Be sure to follow the below instructions when servicing or recycling Well ventilate the room to prevent refrigerant accumulation Keep fire away when the refri...

Страница 31: ...ion und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitdokumenten bedeutet dass gebrauchte elektrische und elektronische Produkte sowie Batterien nicht...

Страница 32: ...ive au produit chimique concern Spanish Eliminaci n de Aparatos Viejos y de Pilas y Bater as Solamente para la Uni n Europea y pa ses con sistemas de reciclado Estos s mbolos en los productos su embal...

Страница 33: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Страница 34: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Страница 35: ...operation the pen point is apart from the recording paper 2 Pull the cartridge out of the mounted position as shown in Fig 1 and Fig 2 3 Set a new recording paper in place on the rear bottom of the c...

Страница 36: ...r of dry cell at the left bottom Fig 3 Set a dry cell in the case with its minus pole positioned backward 4 Close the case cover and replace the cartridge 5 Lift up the pen holder lever and close the...

Страница 37: ...de holes on the pen arm See to Fig 2 illustration 2 5 Press the two sides of the head clamp as indicated by the arrows to secure the ink pen See to Fig 2 illustration 3 From the side view the cartridg...

Страница 38: ...recording paper speed is set to 32 days The center of the recording paper is divided into 32 equal sections The lines extending from these lines serve as the 32 day time scale Fig 4 Stopping recordin...

Страница 39: ...COOLING KIT OPTION Following shows the time to keep chamber temperature at 70o C by using an optional backup cooling kit Keep a liquefied CO2 cylinder at ambient temperature lower than 31o C The above...

Страница 40: ...witch of backup cooling kit BACKUP When turning on the backup cooling kit the backup standby lamp is brightened This means that the backup cooling kit is ready To stop the operation of the backup cool...

Страница 41: ...nner door MDF 7ID is installed Cooling performance 82o C at the center of the chamber ambient temperature 30o C no load An optional inventory rack IR 224U cannot be used when the small inner door is i...

Страница 42: ...ion may cause injury or damage NOTE Only inventory rack IR 220U can be applied when the small inner door MDF 7ID is installed 1 Insert the upper side 2 Push in the lower side Attachment of shelf stopp...

Страница 43: ...High stage side R 290 Low stage side R 170 Temperature controller Microcomputer control system Temperature display Digital display Thermal sensor Platinum resistance Pt 1000 Alarm High temp alarm Low...

Страница 44: ...r of the chamber ambient temperature 30o C no load Temperature control range 50o C to 86o C ambient temperature 30o C no load Power source AC 230 V 240 V 50 Hz AC 230 V 240 V 50 Hz Rated power consump...

Страница 45: ...unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see below Pr...

Страница 46: ......

Страница 47: ......

Страница 48: ...rating Instructions Printed in Japan LDCL031800 0 S0418 0 2018 04 01 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan EU countries only PHC Corporation 2018 Nijverheidsweg 120 4879 AZ Etten Leur...