Страница 1: ...ries Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 39 for all model numbers Operating Instructions Blood Bank R...

Страница 2: ...TERMINAL P 20 TEMPERATURE RECORDER Recorder unit P 21 Loading the pen cartridge P 22 Setting the start time P 23 Replacing the recording chart P 23 CALIBRATION OF SENSOR P 24 ALARM FUNCTIONS P 25 SELF...

Страница 3: ...r MBR 705GR made by PHC Corporation The contents are subject to change without notice No part of the operating instructions may be reproduced in any form without the expressed written permission of PH...

Страница 4: ...action is prohibited This symbol means an instruction must be followed Be sure to keep the operating instructions in a place accessible to users of this unit Label on the unit This mark is labeled on...

Страница 5: ...tall the unit at the location capable of bearing the total combined weight product optional accessories storage items After installing the unit be absolutely sure to take steps to prevent the unit fro...

Страница 6: ...ly onto the unit as this may cause electric shock or short circuit Never put containers with liquid on the unit as this may cause electric shock or short circuit when the liquid is spilled Never disas...

Страница 7: ...ssibly resulting in injury An individual whose job it is to supervise the movement must be put in place and the unit moved while ensuring safety Never replace the battery for power failure alarm Only...

Страница 8: ...g the evaporating pipe during operation of the unit or immediately after stopping of the operation may cause injury Always store the items in the drawers The storage at outside of the drawers may degr...

Страница 9: ...C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltag...

Страница 10: ...10 REFRIGERATOR COMPONENTS Refrigerator unit 1 Fluorescent lamp page 29 Temperature recorder Light switch 3 5 7 Door gasket 2 4 6 9 upper lower 8 both sides top 10 11 Key hole...

Страница 11: ...d in this tray Wipe off the water on the condensation tray with a cloth occasionally 7 Drawer The maximum load is 40 kg per drawer Items to be stored in the chamber must be placed on the drawers Do no...

Страница 12: ...his shows the present chamber temperature and during an alarm status this shows a blinking chamber temperature page 25 And an error code and a chamber temperature is displayed alternately when the sel...

Страница 13: ...l The alarm is noticed to an operator in the remote area by the remote alarm Refer to page 20 for the details 3 Battery switch This is a ON OFF switch for the battery for power failure alarm Always tu...

Страница 14: ...mbient temperature is subject to sudden changes it will not be possible to achieve a stable cooling performance A level location where the floor is capable of bearing the total combined weight product...

Страница 15: ...are raised 5 mm to 10 mm from the surface of the floor Fig 1 In addition rotate the leveling feet slightly clockwise or counterclockwise and adjust them so that the unit is level between the left and...

Страница 16: ...corder sensor as well from the upper monitor bottle 3 Take out the monitor bottle from the fixing plate by lifting up 4 Remove the cap screw type on the monitor bottle as shown in Fig 2 5 Pour 10 glyc...

Страница 17: ...or is closed 3 The chamber temperature is set to 4 o C at the factory Check that the chamber temperature reaches 4 o C on the temperature display Check that the chamber temperature falls to 4 o C on t...

Страница 18: ...he air intake vent They may be vacuumed up and block the air intake vent Air exhaust vent Do not block the air exhaust by the storage items near the air exhaust vent Care should be taken not to put st...

Страница 19: ...temperature display changeover key Lower temperature is displayed UPPER LOWER Lower temperature 4 Press the temperature display changeover key Mean temperature is displayed UPPER LOWER Mean temperatu...

Страница 20: ...is canceled by pressing the alarm buzzer stop key BUZZER since the remote alarm terminal is operated in conjunction with alarm buzzer However the alarm status of remote alarm terminal is not canceled...

Страница 21: ...to turn the power ON and OFF 9 Pilot lamp A green light is seen through the recording chart when the power is ON This lamp will not turn ON when the temperature recorder is being powered by the back...

Страница 22: ...icated by the arrows to open the head clamp at A and B 5 As shown in 2 2 in Fig 2 position the pen cartridge so that the guide pins fit into the guide holes on the pen arm 6 As shown in 2 3 in Fig 2 p...

Страница 23: ...lacing the recording chart 1 Turn off the power switch 2 Slightly raise the end of the pen lifter and remove from the pen lifter stopper Then rotate clockwise until the pen top is on the pen lifter as...

Страница 24: ...mixture allow the temperature indicated on the temperature display to be stable 7 Calibration is needed if the temperature on the temperature display does not show 0 o C 8 Press the calibration key C...

Страница 25: ...However the alarm status of remote alarm terminal is not canceled by pressing the alarm buzzer stop key BUZZER when the unit is under power failure alarm This unit is provided with the alarm buzzer re...

Страница 26: ...disconnected Alarm indicator blinks E17 and chamber temp is displayed alternately If the lower sensor for temp display is short circuited Alarm indicator blinks E18 and chamber temp is displayed alte...

Страница 27: ...t cloth Then wipe off the cabinet or accessories with a dry cloth Important Never pour water onto or into the unit This may cause electric shock or failure CAUTION Do not use a brush an acid a thinner...

Страница 28: ...acket for the evaporating tray and then remove the mounting bracket and the evaporating tray The evaporating pipe is hot Touching the evaporating pipe during operation and immediately after stopping t...

Страница 29: ...he fluorescent lamp downwards with the lamp cover to remove the fluorescent lamp together with the wiring Fig 1 4 After removing the water proof cover remove the socket with the wiring on the both sid...

Страница 30: ...ower switch turn off the light switch if the light switch is on 2 Open the front cover page 13 3 Remove the drip proof cover and the glow starter 4 Install a new glow starter glow starter FG 1P 5 Repl...

Страница 31: ...ction blink of alarm indicator sound of alarm buzzer will not operate when the battery for power failure alarm is flat The alarm indicator blinks and the alarm buzzer sounds by the battery for power f...

Страница 32: ...breaker on the supply circuit is activated A breaker of 15 A or more is recommended The fuse on the supply circuit is blown The outer door was kept opened for a long time The outer door is opened An e...

Страница 33: ...recyclable powers the product you have purchased Please call 1 800 8 BATTERY for information on how to recycle this battery L appareil que vous vous tes procur est aliment par une pile au nickel m ta...

Страница 34: ...the chemical involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitd...

Страница 35: ...r sur les piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci...

Страница 36: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Страница 37: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Страница 38: ...200 W x 1 Evaporator Fin and tube type Condenser Wire and tube type Refrigerant R 134a Defrosting Cycle defrost forced defrost Fully automatic Defrost heater 100 W Temperature controller Electronic c...

Страница 39: ...ribution 1 5 o C Usable ambient temperature 5 o C to 35 o C Noise level 45 dB A scale Maximum pressure 1620 kPa Rated voltage AC 115 V AC 220 V Rated frequency 60 Hz 60 Hz Amperage 3 1 A 1 9 A Power c...