Instructions For Use:
MCO230AICUV - IncuCyte kit
1.
Follow the installation instruction from the original instructions for use of the MCO-230AIC-PA.
2.
Take out the original humidifying pan, humidifying pan cover and the 4 trays (see Fig. 1 and 2).
3.
Place the smaller humidifying pan
with part number 624 230 4548 on
the bottom of the incubator and
make sure to place it with the long
side against the back wall of the
incubator so water accumulated on
the dew stick will drip in this
humidifying pan.
4.
Place the complete humidifying pan cover with part number 624
230 4555 over the humidifying tray and make sure to place it
against the back wall of the incubator.
5.
Place the stand, NU-STAND-230AICGEN, for the IncuCyte S3 system on the bottom of the incubator
6.
Install the IncuCyte S3 inside the incubator. Set temperature to 36.5 °C (see manual of MCO-230AIC-
PA to adjust).
Installation