pH ELECTRODE PE-03 Скачать руководство пользователя страница 5

8)pH readings stabilize faster in some solutions than others;

allow time for the reading to stabilize. In general, buffers
provide stable readings in several seconds (tris buffers
take somewhat longer) while samples usually take longer
times.

9)Keep in mind that all pH electrodes age with time. Aging

is characterized by shortened span and slower speed of
response. If the meter has a manual or microprocessor
slope (span) control, the control can be adjusted to
compensate for electrode span errors ( but will not effect
the speed of response ). Aging is best detected by
calibrating the electrode in, for example, 7 buffer, then
rinsing and placing the electrode in 4 buffer. As a rule, if
the span is 10 % or more in error (a reading of 4.3 or
higher for this example) the electrode should be cleaned and
retested (see the Electrode Cleaning Section) or
reconditioned ( see the Reconditioning Section ). 
If performance is not restored the electrode should be
replaced.

3. CALIBRATION AND MEASUREMENTS

As a rule, follow the procedures recommended by the pH
Meter manufacturer keeping in mind the Helpful Operating
Techniques given above. The frequency of calibration is a
function of the electrode, the pH meter and the solutions
the electro measurements :

1)Remove the electrode form its soaker bottle and save the

bottle.

2)Vigorously stir the electrode in a rinse solution.
3)Shake the electrode with a snap action to remove residual

drops of solution.

4)Vigorously stir the electrode in the buffer or sample and

allow the electrode to rest against the beaker's wall. 

3

Содержание PE-03

Страница 1: ...eld of precision measurement Although this pH ELECTRODE is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please read...

Страница 2: ...TABLE OF CONTENTS 1 SPECIFICATIONS 1 2 HELPFUL OPERATING TECHNIQUES 2 3 CALIBRATION AND MEASUREMENTS 3 4 ELECTRODE STORAGE 4 5 ELECTRODE CLEANING 4 6 RECONDITIONING 5...

Страница 3: ...otential 7 1 pH for pH Value Body Material Epoxy Connector BNC Mechanical With protection bottle on the electrode Protection head Dimensions Body length 160 mm Body Dia 12 mm Cable length 1000 mm Appl...

Страница 4: ...er 3 Vigorously stir the electrode in the sample buffer or rinse solution This action will bring solution to the electrode s surface more quickly and improve speed of response 4 After exposure to a sa...

Страница 5: ...le if the span is 10 or more in error a reading of 4 3 or higher for this example the electrode should be cleaned and retested see the Electrode Cleaning Section or reconditioned see the Reconditionin...

Страница 6: ...ectrode then the o ring then insert the electrode into the bottle and firmly tighten the cap If the solution in the soaker bottle is missing fill the bottle with pH 4 buffer 5 ELECTRODE CLEANING Coati...

Страница 7: ...rse the electrode tip in 0 1N HCL for 15 seconds rinse in tap water and then immerse tip in 0 1N NaOH for is seconds and rinse in tap water Repeat this sequence three times and then recheck the electr...

Отзывы: