Measuring signal (U)
Pressure (p)
Constant (c)
1)
[V]
[hPa]
5.5
[mbar]
[Torr]
5.625
[micron]
2.625
[Pa]
3.5
Tbl. 6:
Constants for converting measuring signal and pressure
Measuring signal (U)
Pressure (p)
[V]
[hPa]
< 0.5
Sensor error
0.5 – 1.5
Underrange
1.5
1.0 × 10
-4
2.5
1.0 × 10
-3
3.5
1.0 × 10
-2
4.5
0.1
5.5
1.0
6.5
10
7.5
1.0 × 10
2
8.5
1.0 × 10
3
8.5 – 10.3
Overrange
Tbl. 7:
Conversion table TPR 270
Measuring signal (U)
Pressure (p)
[V]
[hPa]
< 0.5
Sensor error
0.5 – 2.2
Underrange
2.2
5.0 × 10
-4
2.5
1.0 × 10
-3
3.5
1.0 × 10
-2
4.5
0.1
5.5
1.0
6.5
10
7.5
1.0 × 10
2
8.5
1.0 × 10
3
8.5 – 10.3
Overrange
Tbl. 8:
Conversion table TPR 271
Converting measuring signal and pressure
Validity range
● 1 × 10
-4
hPa < p < 1000 hPa
● 7.5 × 10
-5
Torr < p < 750 Torr
● 1 × 10
-2
Pa < p < 1 × 10
5
Pa
1)
Dependent upon unit of pressure