General notes for the installation of vacuum components
► Choose an installation location that permits access to the product and to supply lines at all times.
► Observe the ambient conditions given for the area of use.
► Provide the highest possible level of cleanliness during assembly.
► Ensure that flange components during installation are grease-free, dust-free and dry.
Procedure
1. Check the carrying capacity of the floor at the installation location.
2. Place the vacuum pump on a flat, horizontal and fixed surface, to safeguard the lubricant supply.
– Reference surface is the vacuum flange.
3. Screw the 4 feet of the vacuum pump on level with the standing surface.
4. If necessary, use adjustment elements from the Pfeiffer Vacuum
the feet horizontally.
5. When doing so, make sure that the vacuum pump is not tensioned.
6. When installing the pump in a closed housing, ensure adequate air circulation.
7. Keep both sight glasses freely accessible for checks and maintenance.
8. Ensure that the motor rating plate remains accessible at all times for a clear view of the voltage
and frequency specifications.
9. Maintain the minimum distances to bordering surfaces to guarantee sufficient air circulation.
10. Leave the filling/drain holes and sight glasses freely accessible.
11. Fill with lubricant prior to first commissioning.
5.3 Filling with lubricant
The lubricant type specified for the roots pump and the respective filling quantity are indicated on the
rating plate. Only the lubricant used during initial installation is permissible. Subsequent change is pos-
sible only after consultation with Pfeiffer Vacuum.
NOTICE
Property damage from using non-approved lubricant
Attainment of product-specific performance data is not ensured. If non-approved lubricants are used,
all liability and warranty claims against Pfeiffer Vacuum are excluded.
► Use approved lubricant only.
► Use alternative, application-specific lubricants only following consultation with Pfeiffer Vacuum.
Required tools
● Open-end wrench,
SW 17
● Calibrated torque wrench (tightening factor ≤ 2.5)
Permissible lubricants
● P3 (standard lubricant)
● D1 for special applications (such as higher operating temperatures)
● Other lubricants on request
Installation
25/60