7.3 Replacing the device fuses
WARNING
Danger to life from electric shock during maintenance and service work
The vacuum pump is only completely de-energized when the mains plug has been disconnected.
There is a danger to life from electric shock when making contact with live components.
► Before performing all work, switch off the mains switch.
► Disconnect mains cable from mains plug.
► Wait 2 minutes until the capacitors have discharged.
► Secure the vacuum pump against unintentional re-start.
The device fuse can be found in a fuse holder on the side housing cover.
Prerequisites
● Supply voltage switched off
● Drive motor disconnected from mains and secured against activation
Required tools
● Screwdriver
Spare part required
● Microfuse 5×20 mm / 2.5 AT
1
2
Fig. 6:
Fuse holder with device fuse
1 Fuse holder
2
Side housing cover
Replacing defective device fuses
1. Open the fuse holder.
2. Replace the defective device fuse.
3. Close fuse holder.
7.4 Change the diaphragms and valves
NOTICE
Property damage from incorrect installation
Change in dead volume due to incorrect installation of the original spacer disks impairs the final vac-
uum or leads to bearing damage.
► During disassembly, keep the existing spacer disks separate per installation location.
► Reinstall the same number of original spacer disks per diaphragm head.
Maintenance
28/46