Pin Assignment
Description, factory setting
1
unassigned
2
DI remote priority
Operation via “remote” interface
3
unassigned
4
DI2
open: off; V+: on
5
DI Pumping station
open: off; V+: on and malfunction acknowledgment
6
DI stand-by
open: Standby off, V+: Standby on
7
+24 V DC* output (V+) Reference voltage for all digital inputs (5 W max.)
8
DO1
GND: no, V+: yes (l
max
= 50 mA/24 V), function can be selected
via parameters
9
DO2
GND: no, V+: yes (l
max
= 50 mA/24 V), function can be selected
via parameters
10
unassigned
11
unassigned
12
unassigned
13
RS-485
D+
14
RS-485
D-
15
Ground (GND)
Reference earth for all digital inputs and outputs
Tbl. 7:
Connection assignment of 15-pin "remote" connection
6.2.1 Inputs
The digital inputs at the "remote" connection are used to switch various electronic drive unit functions.
Inputs are assigned with functions ex-factory. You can configure them via the RS-485 interface and the
Pfeiffer Vacuum parameter set.
DI remote priority/pin 2
V+:
The "remote" connection has control priority over all other control sources.
open:
Remote priority inactive
DI/Pin 4
V+:
On
open:
Off
DI pumping station/pin 5
V+:
Pumping station on and malfunction acknowledgment.
Reset pending error messages after eliminating the cause.
open:
Pumping station off
DI standby/pin 6
V+:
Standby on
open:
Operation at nominal rotation speed or selected rotation speed in rotation speed setting
mode
6.2.2 Outputs
The digital outputs at the "remote" connection have a maximum load limit of 24 V/50 mA per output.
You can configure all listed outputs with the Pfeiffer Vacuum parameter set via the RS-485 interface
(description relates to factory settings).
DO1/Pin 8
Active high
yes
GND
no
Interfaces
33/84