6 Operation
6.1 Commissioning
Important settings and function-related values are programmed into the electronic drive units of the vac-
uum pumps as parameters. Each parameter has a three-digit number and a designation. Parameters
can be used via Pfeiffer Vacuum display and control units or via RS-485 with the Pfeiffer Vacuum proto-
col.
The DCU is used for both parameter sets as a user interface. The HiCube Eco controls two independ-
ent parameter sets for the turbopump and the backing pump. The parameters of the backing pump are
shown in order on the display and are identified with a
symbol.
Parameters Name
Designation
Configuration
HiCube 30 Eco
HiCube 80 Eco
HiCube 300 Eco
[P:001]
Heating
Heating
Without heating
not available
0
0
With heating
1
1
[P:012]
EnableVent
Enable venting
1
1
1
[P:019]
Cfg DO2
Configuration
of output DO2
22 = Backing
pump standby
22 = Backing
pump standby
22 = Backing
pump standby
[P:023]
MotorPump
Motor pump
1
1
1
[P:024]
Cfg DO1
Configuration
of output DO1
15 = Pumping
station
15 = Pumping
station
15 = Pumping
station
[P:025]
Op-
Mode BKP
Backing pump
mode
1 = Intermittent
operation
1 = Intermittent
operation
1 = Intermittent
operation
[P:035]
Cfg Acc A1
Configuration
accessory con-
nection A1
Air cooling
4 = Fan (tem-
perature control-
led)
4 = Fan (tem-
perature con-
trolled)
4 = Fan (temper-
ature controlled)
Water cooling
2 = Heating
2 = Heating
2 = Heating
[P:036]
Cfg Acc B1
Configuration
accessory con-
nection B1
1 = Venting
valve, normally
closed
1 = Venting
valve, normally
closed
1 = Venting
valve, normally
closed
[P:708]
PwrSVal
Set value pow-
er consumption
with MVP 015
‑
2 90 %
90 %
90 %
with MVP 015
‑
4 100 %
100 %
100 %
with MVP 030
100 %
100 %
100 %
[P:710]
Swoff BKP
Switching off
threshold back-
ing pump in in-
termittent mode
25 W
25 W
45 W | MVP 015
52 W | MVP 030
[P:711]
SwOn BKP
Switching on
threshold back-
ing pump in in-
termittent mode
40 W
40 W
55 W | MVP 015
62 W | MVP 030
[P:794]
Param set
Parameter set
1 = extended
1 = extended
1 = extended
[P:795]
Servicelin
Service line
309 = Active ro-
tation speed
309 = Active ro-
tation speed
309 = Active rota-
tion speed
[P: 030]
ValveMode
Configuration
gas ballast
with MVP 015
0 = auto
0 = auto
0 = auto
[P: 717]
StdbySVal
Set value rota-
tion speed in
standby
with MVP 015
66 %
66 %
66 %
Tbl. 6:
Preset setting values for HiCube Eco on delivery
Operation