7.4.2 Draining lubricant on the vacuum side
1
3
2
3
Fig. 14:
Draining lubricant on the vacuum side
1 Magnetic sealing plugs
3
Collection receptacle
2 Drain screw
Procedure
1. Place a collection receptacle beneath the drain hole on the vacuum side.
2. Unscrew the drain screw.
3. Unscrew the magnetic sealing plug.
4. Fully drain the lubricant.
5. Clean the magnetic sealing plug.
6. Screw in the magnetic sealing plug and drain screw.
7.4.3 Filling lubricant on the motor side
MAX
MIN
1
3
2
Fig. 15:
Filling lubricant on the motor side
1 Filler screw
3
Sight glass
2 O-ring
Maintenance
37/52