Fig. 7:
Mains connection
1 Mains cable with C19 bushing from the range of accessories
2
Mains supply plug
Establish the mains supply
1. Order a corresponding mains connection cable from the Pfeiffer Vacuum accessories range.
2. Always ensure a secure connection to the earthed conductor (PE).
3. Plug the mains cable into the power supply plug of the vacuum pump.
On-site protection
The vacuum pump has an internal fuse 10 A TT.
► Recommendation: Always provide a 16 amp fuse as protection for the motor in the event of a mal-
function.
5.6 Installing a residual current circuit breaker
In the event of an insulation fault, installation of a residual current circuit breaker ensures protection
against personal injury.
RCCB
Residual current
waveform
Type B
● AC/DC sensitive
● Safe functioning with overlapping of different residual current waveforms
● regardless of
─ phase angle
─ polarity
─ occurring suddenly or increasing gradually
Tbl. 6:
Technical requirements for a residual current circuit breaker
Procedure
1. Observe the technical requirements for a residual current circuit breaker.
2. Observe the prescribed inspection periods for electrical protective devices.
Installation