Max
.
Min.
1
2
4
3
Fig. 6:
Filling the operating fluid
1 Filler screw with manometer
3
Sight glass
2 O-ring
4
Drain screw
Filling the operating fluid
1. Unscrew the filler screw.
2. Fill the operating fluid according to the sight glass.
3. Screw in the filler screw.
– Be careful with the O-ring.
4. Seal the vacuum connection.
5. Start the vacuum pump.
6. Operate the vacuum pump for approx. 5 minutes.
7. Switch off the vacuum pump.
8. Wait approx. 1 minute until the operating fluid has accumulated in the operating fluid separator.
9. Check the fill level.
10. If necessary, fill more operating fluid in case the fill level is incorrect.
5.5 Connect the exhaust gas temperature monitoring
A thermal circuit breaker on the oil mist filter monitors the gas temperature. The thermal circuit breaker
is preset at the factory.
Installation
26/70