The Pfeiffer Vacuum parameter set
25
6.5
Switching on/off the pump
Switching on
The function "pumping station" comprises turbopump operation with control of all con-
nected accessories (e.g. backing pump).
Switch on the supply voltage with switch S1 on the power supply.
Parameter
[P:023]
= 1 (default)
Parameter
[P:010]
= 1
Ongoing (and removed) error messages are reset. After a successfully completed self-
test, the electronic drive unit sets the turbopump motor and all connected accessories
into operation depending on their configuration.
When the pumping station is activated, the motor of the turbopump can be switched off
and on via the function
[P:023]
.
Switching off
Parameter
[P:010]
= 0
The electronic drive unit switches off the turbopump and activates preset accessory op-
tions (e.g. venting, backing pump).
Wait for the complete standstill of the pump.
Cut off the supply voltage with switch S1 on the power supply.